Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
3.1.2 Span Check
Allow the analyzer to sample span gas through the sample port. Press
CAL
button. After a few
minutes the CO reading should be at the expected concentration. If the correct concentration is
not reached, then the instrument can be adjusted to read the correct value. To do this, press
CONC
and enter the expected concentration of the calibration gas. The
SPAN
button should
now be displayed on the front panel. Pressing
SPAN
and
ENTER
will modify the internals
formulas used to compute the CO reading. If the
SPAN
button is not displayed, this means that
the span reading is too far out of adjustment to do a reliable calibration. The reason for this must
be determined before the analyzer can be calibrated. See Section 10.5 for troubleshooting
calibration problems.
3-2
Summary of Contents for 300M
Page 26: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 1 18 ...
Page 36: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 2 10 ...
Page 56: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 4 10 ...
Page 64: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 5 8 ...
Page 66: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 6 2 ...
Page 86: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 8 8 ...
Page 94: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 9 8 ...
Page 116: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 11 4 ...