Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
Table 5-2: I/O Signals (Continued)
#
Signal
Control
Description
32
DCPS_VOLTAGE
NO
DC power supply composite voltage output. Typically
2500 mV.
33
DAC_CHAN_0
NO
Output of DAC 0(REC) in mV
34
DAC_CHAN_1
NO
Output of DAC 1(DAS) in mV
35
DAC_CHAN_2
NO
Output of DAC 2(TEST) in mV
36
DAC_CHAN_3
NO
Output of DAC 3(Spare) in mV
37
CONC_OUT_1
YES
CO Reading (REC) in mV
38
CONC_OUT_2
YES
CO Reading (DAS) in mV
39
TEST_OUTPUT
YES
Test Channel in mV
40
SAMPLE_LED
YES
Sample LED on/off
41
CAL_LED
YES
Calibration LED on/off
42
FAULT_LED
YES
Fault LED on/off
5.2.2 Analog Output Test
This test cycles the analog output channels from 0% to 100% of full scale in 20% full scale
steps. It starts by outputting 0 volts to all channels and displaying a 0% button. Then, every five
seconds, the output is increased 20% FS and the button is changed accordingly. Thus, the button
(and the analog outputs) will cycle through the following value.
0%, 20%, 40%, 60%, 80%, 100%, 0%, ...
To pause the output at the current voltage, press the
n%
button. To resume automatic cycling,
press the
n%
button again.
5.2.3 Electric Test
This test activates a diagnostic circuit located on the Synchronous Demodulator board which
generates an artificial signal which simulates the output of the IR detector. This signal is injected
in place of the detector output.
When activated, Electric Test will automatically switch the analyzer into a 5000 PPM range and
result in the analyzer producing a constant, stable output (i.e. CO reading) of about 2500 PPM.
This test is particularly useful in isolating problems, since it exercises essentially all electronic
sub-systems of the analyzer but does not depend on the proper function of optical or pneumatic
subsystems.
5-6
Summary of Contents for 300M
Page 26: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 1 18 ...
Page 36: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 2 10 ...
Page 56: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 4 10 ...
Page 64: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 5 8 ...
Page 66: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 6 2 ...
Page 86: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 8 8 ...
Page 94: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 9 8 ...
Page 116: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 11 4 ...