
Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
4.7 Setting the Analog Output Offset
In order to permit the analyzer to connect to a wider variety of strip chart recorders and other
instruments, the analog output of the Carbon Monoxide readings can be adjusted by up to
±
500 mV for 0-5 V range (or
±
10% of current analog output range) in software. The default
output offset is 0 mV. To change it, press
SETUP-MORE-DIAG,
press
NEXT
until D/A
CALIBRATION is displayed and press
ENTR.
Press
CFG
to enter the D/A configuration menu.
Use the
NEXT
and
PREV
buttons to select the desired analog output and press
SET
. Enter a
value of from -500 mV to +500 mV (other ranges will ratio accordingly), followed by
ENTR
to
accept the change, or
EXIT
to leave it unchanged. The offset will be reflected immediately on
the strip chart recorder or other instrument.
4.8 Setting the RS-232 Baud Rate
To set the baud rate for the RS-232 channel, press
SETUP-MORE-COMM-BAUD
. Press
300
,
1200
,
2400
,
4800
,
9600
, or
19.2
followed by
ENTR
to accept the new baud rate, or
EXIT
to
leave the baud rate unchanged.
4.9 Setting the Analyzer I.D.
Each analyzer may be programmed with a unique I.D. number that appears on all RS-232
messages. To set the analyzer I.D., press
SETUP-MORE-COMM-ID
. Enter a 4-digit number
from 0000 to 9999, followed by
ENTR
to accept the new I.D., or
EXIT
to leave the I.D.
unchanged. If changed, the new I.D. number will appear on all RS-232 reports from this
analyzer.
4.10 Disabling the Calibration Password
Normally, operators are required to enter the calibration password when doing a manual
calibration via the
CALZ
,
CALS
, or
CAL
buttons. To allow calibration without entering the
password, press
SETUP-PASS
and set it to OFF, and then press
ENTR
to accept the change, or
EXIT
to leave it unchanged. To enable the calibration password, set the variable to ON.
4-4
Summary of Contents for 300M
Page 26: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 1 18 ...
Page 36: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 2 10 ...
Page 56: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 4 10 ...
Page 64: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 5 8 ...
Page 66: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 6 2 ...
Page 86: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 8 8 ...
Page 94: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 9 8 ...
Page 116: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 11 4 ...