Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
Table 1-2: Final Test and Calibration Values
Test Values
Observed
Value
Units
Nominal Range
CO MEAS
mV
2500 - 4700
CO REF
mV
2500 - 4700
MR RATIO
1.1 – 1.22
SAMPLE PRESS
in-Hg
27 - 30
VACUUM PRESS
in-Hg
10
±
5
SAMPLE FLOW
CC/MIN
700 - 900
BENCH TEMP
°
C 48
±
1
DC POWER
SUPPLY
mV
2500
±
200
SAMPLE TEMP
°
C
45-50
BOX TEMP
°
C
8- 40
WHEEL TEMP
°
C 68
±
2
Diagnostic
Dark Ref
mV
125
±
50
Dark Meas
mV
125
±
50
Electrical test
PPM
400
±
4
Span and CAL Values
CO Span Conc
PPM
400
CO Slope
1.0
±
0.3
CO Offset
-0.1 to +0.1
Noise at Zero (p-p)
PPM
0.5
Noise at Span (p-p)
PPM
<1.0% of reading
Factory Installed Options
Option Installed
Factory Installed Options
Option Installed
Power Voltage/Frequency
Rack Mount, w/ Slides
Internal Zero Span Valves
Analog Voltage Range
0 - V
PROM #
Serial #
Date
Technician
1-16
Summary of Contents for 300M
Page 26: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 1 18 ...
Page 36: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 2 10 ...
Page 56: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 4 10 ...
Page 64: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 5 8 ...
Page 66: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 6 2 ...
Page 86: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 8 8 ...
Page 94: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 9 8 ...
Page 116: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 11 4 ...