Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
Table 7-5: Control Commands
Command Message
Meaning
"C ZERO<CRLF>"
Do a zero check
"C COMPUTE ZERO<CRLF>"
Calibrate Zero point
1
"C SPAN<CRLF>"
Do a span check
"C COMPUTE SPAN<CRLF>"
Calibrate Span point
1
"C ASEQ1<CRLF>"
Do a auto-cal sequence
2
"C ASEQ2<CRLF>"
Do a auto-cal sequence
2
"C ASEQ3<CRLF>"
Do a auto-cal sequence
2
"C ABORT<CRLF>"
Aborts auto-cal sequence
"C EXITZ<CRLF>"
Exit zero cal only
"C EXITS<CRLF>"
Exit span cal only
"C EXIT<CRLF>"
Exit zero span or hold
1
Executed only if the instrument is in the proper calibration mode and concentration is within
calibration limits. This command adjusts slope and offset values.
2
Initiated only If automatic calibration sequence setup is programmed and enabled.
NOTE
The commands in Table 7-5 can only be entered via the RS-232 port
when the analyzer is in the sample mode.
When a control command is issued, the CPU will respond by issuing a status report. For example
if the host computer issues the command.
C CALZ<CRLF>
to do a zero check, the CPU will send the status report
C DDD:HH:MM IIII START ZERO CALIBRATION<CRLF>
to the RS-232 output.
7-7
Summary of Contents for 300M
Page 26: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 1 18 ...
Page 36: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 2 10 ...
Page 56: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 4 10 ...
Page 64: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 5 8 ...
Page 66: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 6 2 ...
Page 86: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 8 8 ...
Page 94: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 9 8 ...
Page 116: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 11 4 ...