SP3X Sensors
LXT-380 Universal Transmitter
Teledyne Analytical Instruments
56
sensor in place. The torque specification for the gland fitting is 20-ft/lbs.
Over-tightening of the nut may swage the nylon or Teflon ferrules to the
housing crushing the internal sensor components.
8.1.2 IMMERSION
The ¾” MNPT compression fitting is reversed and threaded into an
extension/immersion pipe so the compression gland is facing the
measurement end of the sensor. Feed the cable through the immersion
tube, insert the SP3X sensor at least 5” into the tube and tighten the
sensor in place.
8.1.3 FLOW THROUGH
Although the insertion configuration can be used as a flow-through
mounting by inserting the SP3X sensor into a pipe tee, Teledyne has
various flow cells available for convenience. The flow cells are 2.0”
diameter by 5” long, ported ½” FNPT through and ¾” FNPT for the
sensor. The flow cells are available in 316 Stainless Steel, PVC and
Kynar. Use of the flow cell can facilitate an optional spray cleaning
nozzle for the electrode. Connecting 40+ psi of water or air to the nozzle
will remove particulate materials or biofilms from the sensor tip.
Detergents or solvents can be used to remove greases or oils from the
sensor while acids can be used for hard water scale.
8.1.4 VALVE RETRACTABLE
The SP3X sensor is optionally designed for valve retraction service,
(-1) Sensor Style. Mounting is directly into a process line or through a
Summary of Contents for LXT-380
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 19: ...LXT 380 Universal Transmitter Teledyne Analytical Instruments xiv Blank Page...
Page 55: ...Operation LXT 380 Universal Transmitter Teledyne Analytical Instruments 36 Blank Page...
Page 69: ...Maintenance LXT 380 Universal Transmitter Teledyne Analytical Instruments 50 Blank Page...
Page 71: ...Troubleshooting LXT 380 Universal Transmitter Teledyne Analytical Instruments 52 Blank Page...
Page 78: ...LXT 380 Universal Transmitter SP3X Sensors Teledyne Analytical Instruments 59...
Page 91: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 72 B LXT 380 Hart Menu...
Page 92: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 73...
Page 96: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 77...
Page 97: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 78...
Page 99: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 80 Sensor Type...
Page 101: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 82 E Drawings...
Page 102: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 83...
Page 103: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 84...