LXT-380 Universal Transmitter
Calibration
Teledyne Analytical Instruments
41
4.2 ORP CALIBRATION PROCEDURES
AUTO Calibration recognizes Quinhydrone solutions (mVa), pH
7.00 quinhydrone solution (90 mV) and pH 4.01 quinhydrone solution
(267 mV) for automatic ORP calibrations. Any calibration solutions can
be used but the ORP value will have to be entered manually. Follow the
steps below to accomplish an ORP calibration.
4.2.1 AUTO CAL WITH QUINHYDRONE
4.2.2 STANDARDIZE
Leave the sensor in the process solution, take a grab sample from
the process and determine the ORP.
Summary of Contents for LXT-380
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 19: ...LXT 380 Universal Transmitter Teledyne Analytical Instruments xiv Blank Page...
Page 55: ...Operation LXT 380 Universal Transmitter Teledyne Analytical Instruments 36 Blank Page...
Page 69: ...Maintenance LXT 380 Universal Transmitter Teledyne Analytical Instruments 50 Blank Page...
Page 71: ...Troubleshooting LXT 380 Universal Transmitter Teledyne Analytical Instruments 52 Blank Page...
Page 78: ...LXT 380 Universal Transmitter SP3X Sensors Teledyne Analytical Instruments 59...
Page 91: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 72 B LXT 380 Hart Menu...
Page 92: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 73...
Page 96: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 77...
Page 97: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 78...
Page 99: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 80 Sensor Type...
Page 101: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 82 E Drawings...
Page 102: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 83...
Page 103: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 84...