Calibration
LXT-380 Universal Transmitter
Teledyne Analytical Instruments
44
The actual concentration in mg/L (C) is equal to the Saturation
value at the given temperature multiplied by the altitude and air pressure
corrections. Determine the calibration temperature and look up the
saturation value (S) in Table 1 below. Then determine the altitude
correction (K) from Table 2 and the current air pressure in bar (P), 1 bar
equals 14.7 psi. Use 1 bar if the actual air pressure is unknown.
C = S x K x P
Example:
Temperature = 20°C → Saturation = 9.08 mg/L, Altitude = 1200 Ft. →
K = 0.960, Air Pressure 1.014 bar:
C = 9.08 x 0.960 x 1.014 = 8.84 mg/L
The LXT-380 transmitter uses the temperature compensated
Saturation Index for AUTO Cal, however the user can enter the altitude
and pressure compensated value of 8.84 ppm as the calibration value
when prompted to “Accept Value?” in CAL 2.
Table 1 Saturation Index
Summary of Contents for LXT-380
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 19: ...LXT 380 Universal Transmitter Teledyne Analytical Instruments xiv Blank Page...
Page 55: ...Operation LXT 380 Universal Transmitter Teledyne Analytical Instruments 36 Blank Page...
Page 69: ...Maintenance LXT 380 Universal Transmitter Teledyne Analytical Instruments 50 Blank Page...
Page 71: ...Troubleshooting LXT 380 Universal Transmitter Teledyne Analytical Instruments 52 Blank Page...
Page 78: ...LXT 380 Universal Transmitter SP3X Sensors Teledyne Analytical Instruments 59...
Page 91: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 72 B LXT 380 Hart Menu...
Page 92: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 73...
Page 96: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 77...
Page 97: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 78...
Page 99: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 80 Sensor Type...
Page 101: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 82 E Drawings...
Page 102: ...LXT 380 Universal Transmitter Appendix Teledyne Analytical Instruments 83...
Page 103: ...Appendix LXT 380 Universal Transmitter Teledyne Analytical Instruments 84...