Paramagnetic Oxygen Analyzer
Installation
Teledyne Analytical Instruments
25
Figure 3-5: Electrical Connector Panel
For safe connections, ensure that no uninsulated wire extends
outside of the connectors they are attached to. Stripped wire ends must
insert completely into terminal blocks. No uninsulated wiring should be
able to come in contact with fingers, tools or clothing during normal
operation. The instrument must be grounded both internally and
externally as per the Wiring Diagram in the Appendix.
3.3.1 Primary Input Power
The analyzer requires a 115 or 230 VAC, 50 or 60 Hz power source.
The actual input voltage used must show in the window of the
VOLTAGE SELECTOR switch
before
the power source is connected.
See Figure 3-6 for detailed connections.
DANGER: Power is applied to the instrument's circuitry as
long as the instrument is connected to the power
Summary of Contents for 3020 M
Page 2: ......
Page 3: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 7: ......
Page 8: ......
Page 11: ...Model 3020M ATEX Teledyne Analytical Instruments iv Blank Page...
Page 17: ...Model 3020M ATEX Teledyne Analytical Instruments vi Blank Page...
Page 33: ...Introduction Model 3020M ATEX Teledyne Analytical Instruments 10 Blank Page...
Page 43: ...Operational Theory Model 3020M ATEX Teledyne Analytical Instruments 20 Blank Page...
Page 61: ...Installation Model 3020M ATEX Teledyne Analytical Instruments 38 Blank Page...
Page 90: ......
Page 92: ......
Page 94: ......
Page 96: ......
Page 98: ......
Page 100: ......
Page 102: ......
Page 104: ......
Page 105: ...C TELEDYNE...
Page 106: ......
Page 107: ......