Paramagnetic Oxygen Analyzer
Maintenance
Teledyne Analytical Instruments
61
5.4 Fuse Replacement
WARNING: Risk of electric shock high voltage exposed at
the end of enclosure!
The 3020M requires two 5 x 20 mm, 4 A, T type (Slow Blow)
fuses. The fuses are located inside the explosion proof housing on the
Electrical Connector Panel, as shown in Figure 5-2. To replace a fuse:
1.
Disconnect the Unit from its power source.
2.
Place a small screwdriver in the notch in the fuse holder cap,
push in, and rotate 1/4 turn. The cap will pop out a few
millimeters. Pull out the fuse holder cap and fuse, as shown in
Figure 5-3.
3.
Replace fuse by reversing process in step 1.
Figure 5-2: Removing Fuse Cap and Fuse from Holder
Summary of Contents for 3020 M
Page 2: ......
Page 3: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 7: ......
Page 8: ......
Page 11: ...Model 3020M ATEX Teledyne Analytical Instruments iv Blank Page...
Page 17: ...Model 3020M ATEX Teledyne Analytical Instruments vi Blank Page...
Page 33: ...Introduction Model 3020M ATEX Teledyne Analytical Instruments 10 Blank Page...
Page 43: ...Operational Theory Model 3020M ATEX Teledyne Analytical Instruments 20 Blank Page...
Page 61: ...Installation Model 3020M ATEX Teledyne Analytical Instruments 38 Blank Page...
Page 90: ......
Page 92: ......
Page 94: ......
Page 96: ......
Page 98: ......
Page 100: ......
Page 102: ......
Page 104: ......
Page 105: ...C TELEDYNE...
Page 106: ......
Page 107: ......