Paramagnetic Oxygen Analyzer
Operational Theory
Teledyne Analytical Instruments
17
The sample or calibration gas flowing through the system is moni-
tored by a flowmeter downstream from the sensor.
The gases delivered to the instrument should be at constant
pressures and flow rates and must exit freely into the ambient
atmosphere. The Span, Zero and Sample gases should be delivered at
constant pressures of about 10 psig (Range 5-25 psig). The flow rate
must be maintained at about 300 cc/minute (Range 50-600cc/m) and
must exit freely into atmospheric pressure.
Figure 2-3 is the flow diagram for the sampling system. In the
standard instrument (shown on the right), calibration gases (zero and
span) can be connected directly to the Sample In port by teeing to the
port with appropriate valves. The shaded portions of the diagram show
the components added when the —C and/or —F options are ordered.
The solenoid valves, when supplied, are installed inside the 3020M
enclosure and are regulated by the instruments internal electronics. The
flame arrestors, when supplied, are installed in the Gas Connector
Panel..
Figure 2-3: Flow Diagram
Summary of Contents for 3020 M
Page 2: ......
Page 3: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 7: ......
Page 8: ......
Page 11: ...Model 3020M ATEX Teledyne Analytical Instruments iv Blank Page...
Page 17: ...Model 3020M ATEX Teledyne Analytical Instruments vi Blank Page...
Page 33: ...Introduction Model 3020M ATEX Teledyne Analytical Instruments 10 Blank Page...
Page 43: ...Operational Theory Model 3020M ATEX Teledyne Analytical Instruments 20 Blank Page...
Page 61: ...Installation Model 3020M ATEX Teledyne Analytical Instruments 38 Blank Page...
Page 90: ......
Page 92: ......
Page 94: ......
Page 96: ......
Page 98: ......
Page 100: ......
Page 102: ......
Page 104: ......
Page 105: ...C TELEDYNE...
Page 106: ......
Page 107: ......