Paramagnetic Oxygen Analyzer
Operation
Teledyne Analytical Instruments
43
4.2.2.2
ESCAPE
A turn-and-release toward ESCAPE moves the blinking to the next
field on the left. When you are on the leftmost field, another ESCAPE
takes you back to the previous screen.
If you do not wish to continue a function, you can abort the session
by escaping to the leftmost field, and then issuing another ESCAPE.
Escaping a function takes the analyzer back to the previous screen, or to
the ANALYZE Function, depending on the function escaped.
4.3 The AUTO-CAL Function
When proper automatic valving is connected (see chapter 3,
installation
), the Analyzer can cycle itself through a sequence of steps
that automatically zero and span the instrument.
Note: If you require highly accurate timing of your
AUTO-CAL
,
use external
AUTO-CAL
control where possible. The
internal clock in the Model 3020M is accurate to 2-3 %.
Accordingly, internally scheduled calibrations can vary 2-
3 % per day.
To setup an
AUTO-CAL
cycle:
Scroll to
AUTO-CAL
, and ENTER. A new screen for
Span/Zero
set
appears.
Span OFF Nxt: 0d 0h
Zero OFF Nxt: 0d 0h
If
SPAN
(or
ZERO
) is not flashing, scroll with DOWN/UP control
to
SPAN
(or
ZERO
), then ENTER.
(You won’t be able to set
OFF
to
ON
if a zero interval is entered.)
A
Span Every
... (or
Zero Every
...) screen
appears.
Span Every 0 d
Start 0 h from now
Use DOWN/UP control to set an interval value, and ENTER. Then
use DOWN/UP to set a start-time value, and ENTER.
Span OFF Nxt: 0d 0h
Zero OFF Nxt: 0d 0h
When the
Span/Zero
values screen appears, use the scroll DOWN to
blink the
OFF
/
ON
field
of the
SPAN
(or
ZERO
) function. Use
Summary of Contents for 3020 M
Page 2: ......
Page 3: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 7: ......
Page 8: ......
Page 11: ...Model 3020M ATEX Teledyne Analytical Instruments iv Blank Page...
Page 17: ...Model 3020M ATEX Teledyne Analytical Instruments vi Blank Page...
Page 33: ...Introduction Model 3020M ATEX Teledyne Analytical Instruments 10 Blank Page...
Page 43: ...Operational Theory Model 3020M ATEX Teledyne Analytical Instruments 20 Blank Page...
Page 61: ...Installation Model 3020M ATEX Teledyne Analytical Instruments 38 Blank Page...
Page 90: ......
Page 92: ......
Page 94: ......
Page 96: ......
Page 98: ......
Page 100: ......
Page 102: ......
Page 104: ......
Page 105: ...C TELEDYNE...
Page 106: ......
Page 107: ......