Performance Tests
4- 16
TDS 420A, TDS 430A & TDS 460A Service Manual
H
The digitizing oscilloscope must have been last adjusted at an ambient
temperature b20
_
C and +30
_
C, must have been operating for a
warm-up period of at least 20 minutes, and must be operating at an ambient
temperature between 5
_
C and +40
_
C. (The warm-up requirement is usually
met in the course of meeting the prerequisites listed above.)
Equipment Required
These procedures use external, traceable signal sources to directly check
warranted characteristics. The test equipment required is shown in Table 4--1.
Table 4- 1: Test Equipment
Item Number and
Description
Minimum Requirements
Example
Purpose
1.
Attenuator,10X
(three required)
Ratio: 10X; impedance 50
Ω
; connec-
tors: female BNC input, male BNC
output
Tektronix part number
011-0059-02
Signal Attenuation
2.
Attenuator, 5X
(two required)
Ratio: 5X; impedance 50
Ω
; connec-
tors: female BNC input, male BNC
output
Tektronix part number
011-0060-02
Signal Attenuation
3.
Termination, 50
Ω
Impedance 50
Ω
; connectors: female
BNC input, male BNC output
Tektronix part number
011-0049-01
Checking delay match be-
tween channels
4.
Termination, 75
Ω
Impedance 75
Ω
; connectors: female
BNC input, male BNC output
Tektronix part number
011-0102-01
Used to test Video Option 05
equipped oscilloscopes only
5.
Cable, Precision Coaxial
(three required)
50
Ω
, 36 in, male to male BNC
connectors
Tektronix part number
012-0482-00
Signal Interconnection
6.
Cable, Coaxial
75
Ω
, 36 in, male to male BNC
connectors
Tektronix part number
012-1338-00
Used to test Video Option 05
equipped oscilloscopes only
7.
Connector, Dual-Banana
(two required)
Female BNC to dual banana
Tektronix part number
103-0090-00
Various Accuracy Tests
8.
Connector, BNC “T”
Male BNC to dual female BNC
Tektronix part number
103-0030-00
Checking Trigger Sensitivity
9.
Coupler, Dual-Input
Female BNC to dual male BNC
Tektronix part number
067-0525-02
Checking Delay Match Be-
tween Channels
10. Generator, DC Calibra-
tion
Variable amplitude to
±
110 V; accura-
cy to 0.1%
Wavetek 9100 (or, optionally,
Data Precision 8200, with
1 kV option installed)
Checking DC Offset and
Measurement Accuracy
11. Generator, Sine Wave
100 kHz to at least 400 MHz. Variable
amplitude from 12 mV to 2 V
p-p
.
Frequency accuracy >2.0%
Rohde & Schwarz SMY
1
Checking Analog Bandwidth,
Trigger Sensitivity, Sample-
rate, External Clock, and
Delay-Time Accuracy
12. Meter, Level and Power
Sensor
Frequency range:10 MHz to 400MHz.
Amplitude range: 6 mVp-p to 2 V
p-p
Rohde & Schwarz URV 35,
with NRV-Z8 power sensor
Checking Analog Bandwidth
and Trigger Sensitivity
Summary of Contents for TDS 420A
Page 16: ...Service Safety Summary xii TDS 420A TDS 430A TDS 460A Service Manual ...
Page 36: ...Nominal Traits 1 14 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 44: ...Warranted Characteristics 1 22 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 54: ...Installation 2 6 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 64: ...Operating Information 2 16 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 72: ...Performance Verification Procedures 4 4 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 160: ...Maintenance 6 8 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 226: ...Removal and Installation Procedures 6 74 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 260: ...Options and Accessories 7 6 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 262: ...Electrical Parts List 8 2 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 272: ...Diagrams 9 10 TDS 420A TDS 430A TDS 460A Service Manual ...