Adjustment Procedures
5- 16
TDS 420A, TDS 430A & TDS 460A Service Manual
Use a small standard screw driver to
pry between the cover and metal cord
connector to pop off cover.
Repeat for lower cover.
Figure 5- 6: Exposing the Probe Body
b.
Hook up test-signal source:
H
Connect the high-amplitude output of a pulse generator, through a
50
Ω
precision cable to the input of the pulse generator head (see
Figure 5--7).
H
Connect the output of the pulse generator head through a 10X
attenuator to CH 1.
H
Set the triggering level of the pulse generator head to minimum.
Pulse Generator
Digitizing Oscilloscope
Pulse Generator
Head
10X Attenuator
Precision Cable
Output
Figure 5- 7: Initial Test Hookup
c.
Initialize the oscilloscope:
H
Press save/recall
SETUP
.
Summary of Contents for TDS 420A
Page 16: ...Service Safety Summary xii TDS 420A TDS 430A TDS 460A Service Manual ...
Page 36: ...Nominal Traits 1 14 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 44: ...Warranted Characteristics 1 22 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 54: ...Installation 2 6 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 64: ...Operating Information 2 16 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 72: ...Performance Verification Procedures 4 4 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 160: ...Maintenance 6 8 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 226: ...Removal and Installation Procedures 6 74 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 260: ...Options and Accessories 7 6 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 262: ...Electrical Parts List 8 2 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 272: ...Diagrams 9 10 TDS 420A TDS 430A TDS 460A Service Manual ...