Brief Procedures
4- 6
TDS 420A, TDS 430A & TDS 460A Service Manual
e.
Confirm the four adjustment sections have passed status:
H
Press
SHIFT
; then press
UTILITY
.
H
Press the main menu button
System
until
Cal
is highlighted in the
pop-up menu.
H
Verify that the word
Pass
appears in the main menu under the
following menu labels:
Voltage Reference
,
High Frequency
Response
,
Low Frequency Response
, and
Trigger Skew
. (See
Figure 4--2.)
Verify Pass status for the
adjustment sections.
Run a signal path
compensation and verify
status is Pass.
Display the
CAL
menu.
1
2
3
Figure 4- 2: Verifying Adjustments and Signal Path Compensation
This oscilloscope lets you compensate the internal signal path used to acquire the
waveforms you acquire and measure. By executing the signal path compensation
feature (SPC), you optimize the oscilloscope capability to make accurate
measurements based on the ambient temperature.
Summary of Contents for TDS 420A
Page 16: ...Service Safety Summary xii TDS 420A TDS 430A TDS 460A Service Manual ...
Page 36: ...Nominal Traits 1 14 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 44: ...Warranted Characteristics 1 22 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 54: ...Installation 2 6 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 64: ...Operating Information 2 16 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 72: ...Performance Verification Procedures 4 4 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 160: ...Maintenance 6 8 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 226: ...Removal and Installation Procedures 6 74 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 260: ...Options and Accessories 7 6 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 262: ...Electrical Parts List 8 2 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 272: ...Diagrams 9 10 TDS 420A TDS 430A TDS 460A Service Manual ...