Adjustment Procedures
TDS 420A, TDS 430A & TDS 460A Service Manual
5- 13
H
Plug the probe tip from the probe on CH 1 into the BNC to probe tip
adapter.
Pull forward to
remove the
retractable hook tip.
Unscrew the ribbed
ferrule tip cover and pull
forward to remove.
Pull forward to
remove the
ground-lead
assembly.
3
2
1
Figure 5- 4: Exposing the Inner Probe Tip
Leveled Sign
Wave Generator
Digitizing Oscilloscope
BNC Female to
BNC Female
Adapter
BNC to Probe
Tip Adapter
Output
Figure 5- 5: Initial Test Hookup
c.
Initialize the oscilloscope:
H
Press save/recall
SETUP
.
H
Press the main-menu button
Recall Factory Setup
.
H
Press the side-menu button
OK Confirm Factory Init
.
Summary of Contents for TDS 420A
Page 16: ...Service Safety Summary xii TDS 420A TDS 430A TDS 460A Service Manual ...
Page 36: ...Nominal Traits 1 14 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 44: ...Warranted Characteristics 1 22 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 54: ...Installation 2 6 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 64: ...Operating Information 2 16 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 72: ...Performance Verification Procedures 4 4 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 160: ...Maintenance 6 8 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 226: ...Removal and Installation Procedures 6 74 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 260: ...Options and Accessories 7 6 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 262: ...Electrical Parts List 8 2 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 272: ...Diagrams 9 10 TDS 420A TDS 430A TDS 460A Service Manual ...