Warranted Characteristics
1- 20
TDS 420A, TDS 430A & TDS 460A Service Manual
Table 1- 16: Certifications and compliances
Category
Standards or description
EC Declaration of Conformity --
EMC
1
Meets intent of Directive 89/336/EEC for Electromagnetic Compatibility. Compliance was
demonstrated to the following specifications as listed in the Official Journal of the European
Communities:
EN 61326
EMC requirements for Class A electrical equipment for
measurement, control and laboratory use.
2
IEC 61000--4--2
Electrostatic discharge immunity (Performance criterion B)
IEC 61000--4--2
Electrostatic discharge immunity (Performance criterion B)
IEC 61000--4--3
RF electromagnetic field immunity
3
(Performance criterion A)
IEC 61000--4--4
Electrical fast transient / burst immunity (Performance criterion B)
IEC 61000--4--5
Power line surge immunity (Performance criterion B)
IEC 61000--4--6
Conducted RF immunity (Performance criterion A)
IEC 61000--4--11
Voltage dips and interruptions immunity (Performance criterion B)
EN 61000--3--2
AC power line harmonic emissions
Australia / New Zealand
Declaration of Conformity
Complies with EMC provision of Radiocommunications Act per the following standard(s):
Declaration of Conformity --
EMC
AS/NZS 2064.1/2
Industrial, Scientific, and Medical Equipment: 1992
FCC Compliance
Emissions comply with FCC Code of Federal Regulations 47, Part 15, Subpart B, Class A Limits.
1
To maintain emission requirements when connecting cables to this oscilloscope, use only a high-quality, double-shielded
(braid and foil) cable. The cable shield must have low impedance connections to both connector housings. Acceptable
cables are listed:
GPIB
Tektronix part numbers 012-0991-00, -01, -02 and -03
Centronics
Tektronix part number 012-1214-00
RS-232
CA part number 0294-9
VGA Video
LCOM part number CTL3VGAMM-5
2
Emissions which exceed the levels required by this standard may occur when this equipment is connected to a test
object.
3
≤
0.5 division waveform displacement, or
≤
1.0 division increase in peak-to-peak noise when subjected to a 3 V/m
electromagnetic field over the frequency range of 80 MHz to 1000 MHz.
Summary of Contents for TDS 420A
Page 16: ...Service Safety Summary xii TDS 420A TDS 430A TDS 460A Service Manual ...
Page 36: ...Nominal Traits 1 14 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 44: ...Warranted Characteristics 1 22 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 54: ...Installation 2 6 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 64: ...Operating Information 2 16 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 72: ...Performance Verification Procedures 4 4 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 160: ...Maintenance 6 8 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 226: ...Removal and Installation Procedures 6 74 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 260: ...Options and Accessories 7 6 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 262: ...Electrical Parts List 8 2 TDS 420A TDS 430A TDS 460A Service Manual ...
Page 272: ...Diagrams 9 10 TDS 420A TDS 430A TDS 460A Service Manual ...