Performance Tests
3- 30
RSA3408B Analyzer Service Manual
3.
Prepare for the test:
a.
Hook up the instruments:
Connect the signal generator output through a 50
Ω
N-N coaxial cable to
the RSA3408B
Input
. See Figure 3--10.
50
Ω
N-N coaxial cable
Output
Signal generator
RSA3408B
Figure 3- 10: Hookup for checking the absolute amplitude accuracy
b.
Initialize the analyzer:
H
Press the
System
key on the front panel.
H
Press the
Reset All to Factory Defaults
side key.
c.
Modify the default settings:
H
Press the
Frequency/Channel
key on the front panel.
H
The
Center Freq
menu item is selected by default.
Set the frequency to 25 MHz using the numeric keypad.
H
Press the
Span
key on the front panel.
H
The
Span
menu item is selected by default.
Set the span to 10 MHz using the general purpose knob.
d.
Calibrate the analyzer gain:
H
Press
Cal
>
Calibrate Gain
.
H
Press
Amplitude
>
Auto Level
.
Summary of Contents for RSA3408
Page 4: ......
Page 14: ...Service Safety Summary x RSA3408B Analyzer Service Manual...
Page 18: ...Introduction xiv RSA3408B Analyzer Service Manual...
Page 20: ...Specifications 1 2 RSA3408B Analyzer Service Manual...
Page 26: ...Theory of Operation 2 6 RSA3408B Analyzer Service Manual...
Page 82: ...Performance Tests 3 56 RSA3408B Analyzer Service Manual...
Page 200: ...Diagrams 6 2 RSA3408B Analyzer Service Manual...
Page 202: ...6 4 RSA3408B Analyzer Service Manual...
Page 204: ...6 6 RSA3408B Analyzer Service Manual...
Page 210: ......
Page 211: ......
Page 235: ......
Page 236: ......