Removal and Installation Procedures
5- 54
RSA3408B Analyzer Service Manual
5.
Remove the bandpass filter (YTF):
a.
Using an SMA torque driver, disconnect the SMA cables W106 and
W123 from the 3 dB attenuators at both sides of the bandpass filter.
b.
Using a
5
@
16
inch wrench, disconnect the 3 dB attenuators from the filter.
c.
Unplug the cable W115 from the A150 YTF interface board.
d.
Using a #1 Phillips screwdriver, remove the two screws securing the
bandpass filter to the YTF plate. Lift up the filter with the plate.
e.
Using a No. 2 Pozidriv screwdriver, remove the four screws securing the
bandpass filter to the YTF plate.
f.
Unsolder the A150 YTF interface board at four points to remove it from
the bandpass filter.
W106
W123
When removing the A150 YTF interface
board, unsolder it at four points.
W114A
W114
W124
W121
W123
W119
W120
W101
W115
Band relay
Bandpass filter (YTF)
Input relay
Figure 5- 27: RF1 modules removal - 2