Performance Tests
3- 16
RSA3408B Analyzer Service Manual
Figure 3- 7: Performance test result display
6.
Disconnect the test equipment:
a.
Disconnect the cable from the analyzer input.
b.
Turn off the analyzer and generators.
c.
Disconnect the GPIB cables from all the ports.
Performs the auto tests using two signal generators, Item 1 that covers the
baseband and RF1 band and Item 2 that covers the RF2 and RF3 bands.
Equipment
Required
Two signal generators (Items 1 and 2)
One 50
Ω
N-N coaxial cable (Item 11)
One 10 dB attenuator (Item 14)
One GPIB cable (Item 13, two required)
One SG flatness floppy disk (Item 16, two required)
Prerequisites
Refer to page 3--10
Procedure 2
Summary of Contents for RSA3408