Adjustment Procedures
RSA3408B Analyzer Service Manual
4- 9
Power reference output
Power meter
Figure 4- 3: Hookup for calibrating the power sensor
b.
Press the
Sensor Cal
side key to perform zero adjustment and calibra-
tion for the sensor.
5.
Perform the flatness measurement for the generator:
a.
Modify the hookup:
Connect the signal generator output through an N-N
coaxial cable followed by a 10 dB attenuator and an N adapter to the
power sensor input. See Figure 4--2 on page 4--7 again.
b.
Insert a floppy disk to the disk drive of the analyzer for storing the
flatness data.
c.
Press the
Macro Setup
key on the front panel.
d.
Press
Go to page 2
>
Create SG Flatness...
>
All
using the side key.
When the measurement has been completed, the flatness data is saved to the
file(s) on the floppy disk:
sgrf1flatness.cal (for the baseband and RF1 band)
sgrf2flatness.cal (for the RF2 band)
sgrf3flatness.cal (for the RF3 band)
6.
Disconnect the hookup:
a.
Remove the floppy disk by pressing the eject button.
b.
Turn off the signal generator, power meter, and analyzer.
c.
Disconnect the GPIB cables.
d.
Disconnect the power sensor and the N-N cable.
Summary of Contents for RSA3408
Page 4: ......
Page 14: ...Service Safety Summary x RSA3408B Analyzer Service Manual...
Page 18: ...Introduction xiv RSA3408B Analyzer Service Manual...
Page 20: ...Specifications 1 2 RSA3408B Analyzer Service Manual...
Page 26: ...Theory of Operation 2 6 RSA3408B Analyzer Service Manual...
Page 82: ...Performance Tests 3 56 RSA3408B Analyzer Service Manual...
Page 200: ...Diagrams 6 2 RSA3408B Analyzer Service Manual...
Page 202: ...6 4 RSA3408B Analyzer Service Manual...
Page 204: ...6 6 RSA3408B Analyzer Service Manual...
Page 210: ......
Page 211: ......
Page 235: ......
Page 236: ......