Troubleshooting
5- 80
RSA3408B Analyzer Service Manual
Required Adjustments After Replacing Modules
When you change the following parts or modules, readjustment and/or recalibra-
tion is necessary. For details, refer to
Adjustment Procedures
.
RF3 Module
H
You must change the gain and offset values of YTO Tune DAC in the
DEBUG menu. You can find the gain and offset values in a label located on
the RF3 module.
H
You must calibrate IF flatness, RF gain and flatness for RF1, 2, and 3 bands.
RF2 and RF5 Modules
H
You must calibrate IF flatness, RF gain and flatness for RF1, 2, and 3 bands.
A10 A/D Board
H
You must calibrate baseband gain and baseband flatness.
H
You must calibrate IF flatness and RF gain for RF1, 2, and 3 bands.
H
You must calibrate IQ gain.
Step Attenuator
H
You must calibrate baseband gain and flatness.
H
You must calibrate RF gain and RF flatness for RF1, 2, and 3 bands.
Bandpass Filter (YTF)
H
You must change the gain and offset values of YTF Tune DAC in the
DEBUG menu. You can find the gain and offset values in a label located on
the YTF module.
H
You must calibrate IF flatness, RF gain and flatness for RF2 and 3 bands.
A100 Cal Generator Board
H
You must calibrate baseband gain and RF gain for RF1 band.
Hard Disk Drive
H
Full calibration is necessary because the calibration factors are lost.
Summary of Contents for RSA3408
Page 4: ......
Page 14: ...Service Safety Summary x RSA3408B Analyzer Service Manual...
Page 18: ...Introduction xiv RSA3408B Analyzer Service Manual...
Page 20: ...Specifications 1 2 RSA3408B Analyzer Service Manual...
Page 26: ...Theory of Operation 2 6 RSA3408B Analyzer Service Manual...
Page 82: ...Performance Tests 3 56 RSA3408B Analyzer Service Manual...
Page 200: ...Diagrams 6 2 RSA3408B Analyzer Service Manual...
Page 202: ...6 4 RSA3408B Analyzer Service Manual...
Page 204: ...6 6 RSA3408B Analyzer Service Manual...
Page 210: ......
Page 211: ......
Page 235: ......
Page 236: ......