Performance Tests
3- 22
RSA3408B Analyzer Service Manual
2.
Check frequency accuracy in the baseband:
a.
Modify the analyzer default settings:
H
Press the
Frequency/Channel
key on the front panel.
H
The
Center Freq
menu item is selected by default.
Set the frequency to 10 MHz using the numeric keypad.
H
Press the
Span
key on the front panel.
H
The
Span
menu item is selected by default.
Set the span to 1 MHz using the general purpose knob.
b.
Measure peak frequency:
H
Press the
Peak
key to place the marker on the signal peak.
H
Read the marker readout. Confirm that the value is
10 MHz
1 kHz.
3.
Check frequency accuracy in the RF1 band:
a.
Modify the generator control:
Frequency
2 GHz
. . . . . . . . . . . . . . . .
b.
Modify the analyzer controls:
H
Press the
Frequency/Channel
key on the front panel.
H
The
Center Freq
menu item is selected by default.
Set the frequency to 2 GHz using the numeric keypad.
c.
Measure peak frequency:
H
Press the
Peak
key to place the marker on the signal peak.
H
Read the marker readout. Confirm that the value is
2 GHz
1.4 kHz.
Summary of Contents for RSA3408
Page 4: ......
Page 14: ...Service Safety Summary x RSA3408B Analyzer Service Manual...
Page 18: ...Introduction xiv RSA3408B Analyzer Service Manual...
Page 20: ...Specifications 1 2 RSA3408B Analyzer Service Manual...
Page 26: ...Theory of Operation 2 6 RSA3408B Analyzer Service Manual...
Page 82: ...Performance Tests 3 56 RSA3408B Analyzer Service Manual...
Page 200: ...Diagrams 6 2 RSA3408B Analyzer Service Manual...
Page 202: ...6 4 RSA3408B Analyzer Service Manual...
Page 204: ...6 6 RSA3408B Analyzer Service Manual...
Page 210: ......
Page 211: ......
Page 235: ......
Page 236: ......