Load models
Load models are important electrical considerations when working with the
P6910 probe.
The low frequency model is typically adequate for rise and fall times of 1 ns
or greater in a typical 25
Ω
source impedance environment (50
Ω
runs with a
pass-through connection). For source impedance outside this range, and/or rise
and fall times less than 1 ns, use the high frequency model to determine if a
signi
fi
cant difference is obtained in the modeling result.
The following electrical model displays a single podlet load model of the probe.
(See Figure 14.)
Figure 14: Single podlet load model
Speci
fi
cations
The following tables list the electrical and environmental speci
fi
cations for the
P6910 probe.
The electrical speci
fi
cations apply when the probe is connected between a
compatible logic analyzer and a SUT. Refer to the
Tektronix Logic Analyzer Series
Product Speci
fi
cations & Performance Veri
fi
cation Technical Reference Manual
(available on the TLA Documentation CD or downloadable from the Tektronix
Web site) for a complete list of system and module speci
fi
cations. The probes are
designed to meet Tektronix standard 062-2847-00 class 5.
Table 7: Mechanical and electrical speci
fi
cations
Characteristic
Description
Threshold accuracy
±(35 mV ±1% of setting)
Input resistance
20 k
Ω
±1%
Input capacitance
<1.0 pF
1
Minimum digital signal swing
300 mV single-ended
Maximum nondestructive input signal to
probe
±15 V
Delay from probe tip to input connector
7.70 ns ±80 ps
24
P6910 General Purpose Probe Instruction Manual
Summary of Contents for P6910
Page 4: ......
Page 12: ...Compliance information viii P6910 General Purpose Probe Instruction Manual...
Page 14: ...Preface x P6910 General Purpose Probe Instruction Manual...
Page 40: ...Reference 26 P6910 General Purpose Probe Instruction Manual...
Page 44: ...Replaceable parts 30 P6910 General Purpose Probe Instruction Manual...
Page 46: ...Glossary 32 P6910 General Purpose Probe Instruction Manual...