Specifications
OI1125 E/O Transmitter Instruction Manual
23
Table 8: Electromagnetic compatibility (EMC) (cont.)
Specification
Description
Immunity, conducted disturbances induced
by RF fields
No instrument failures when the instrument power leads are injected with a 150 kHz to
80 MHz, 3 V
RMS
signal 80% amplitude modulated at 1 kHz.
IEC 61000-4-6
1
2/50 us Tr/Th voltage into open circuit, 8/20 us Tr/Th current into short circuit.
Table 9: Safety characteristics
Specification
Description
Power consumption
<40 W when powered from AC line
Source frequency
47 to 63 Hz
Safety certifications
UL3111-1, CAN/CSA-C22.2 No. 1010.1
EC Declaration of Conformity - Low Voltage
Compliance was demonstrated to the following specification as listed in the Official
Journal of the European Communities:
Low Voltage Directive 73/23/EEC, amended by 93/68/EEC
EN 61010-1/A2:1995
Safety requirements for electrical equipment for measure-
ment control and laboratory use.
Laser Safety Label
The label shown in Figure 10 is affixed to the top of the instrument cabinet. To
prevent personal injury, it is important that you heed all cautions when working
with laser equipment.
INVISIBLE LASER RADIATION
AVOID DIRECT
EXPOSURE TO BEAM
MAXIMUM 20mW at 1550nm
CLASS IIIb LASER PRODUCT
INVISIBLE LASER RADIATION
DO NOT STARE INTO BEAM
INSTRUMENTS
CLASS 3A LASER PRODUCT
MAXIMUM 20mW at 1550nm
IEC 60825-1:1993+A1:1997
CAUTION
Figure 10: Laser warning label
Summary of Contents for OI1125
Page 4: ......
Page 8: ...Table of Contents iv OI1125 E O Transmitter Instruction Manual...
Page 20: ...Getting Started 8 OI1125 E O Transmitter Instruction Manual...
Page 30: ...Operating Basics 18 OI1125 E O Transmitter Instruction Manual...
Page 36: ...Specifications 24 OI1125 E O Transmitter Instruction Manual...
Page 38: ......
Page 42: ...Theory of Operation 30 OI1125 E O Transmitter Instruction Manual...
Page 72: ...Performance Verification 60 OI1125 E O Transmitter Instruction Manual...