Specifications
OI1125 E/O Transmitter Instruction Manual
21
Table 7: Environmental characteristics
Specification
Description
Temperature, nonoperating
(–55
_
C to +75
_
C)
The environmental exposure is the procedure stated in Tektronix Design Standard
062-2847-00 for Class 6 equipment.
Temperature, operating
0
_
C to +40
_
C (32
_
F to 104
_
F)
This procedure conforms to the temperature portion of the test in MIL-T-28800E for
Type III, Class 6 equipment, using the limits above.
Humidity, nonoperating
0% to 75% relative humidity from +30
_
C to +60
_
C.
Humidity, operating
0% to 80% relative humidity to +31
_
C, decreasing at a linear rate to 50% at +40
_
C,
minimum.
Random vibration, nonoperating
2.46 g
RMS
, from 5 to 500 Hz, 10 minutes each axis.
Tektronix Standard 062-2858-00 rev. B modified, Random Vibration, Tektronix Class 3
with power spectral density break-points as follows:
Constant 0.020 g
2
/Hz from 5 Hz to 100 Hz, then falls at –3 dB/octave from 100 to
200 Hz, then constant 0.010 g
2
/Hz from 200 to 350 Hz, then falls at –3 dB/octave from
350 to 500 Hz to 0.007 g
2
/Hz @ 500 Hz.
Random vibration, operating
0.31 g
RMS
, from 5 to 500 Hz, 10 minutes each axis.
Tektronix Standard 062-2858-00 rev. B, Random Vibration, Tektronix Class 3
Packaged product vibration and shock
The packaged product qualifies under the Distribution Cycle 1 Assurance Level II for
packaged products 0 to 20 lbs. Test 2 for Warehouse and Vehicle Stacking (compression)
is omitted.
Tektronix standard 062-2858-00, Rev. B.
Altitude, operating and nonoperating
Operating:
3000 m (9847 ft)
Nonoperating:
12190 m (40,000 ft)
MIL-T-2880E modified for more exposure time and more severe test levels than required
for any Type III equipment regardless of Class.
Weight
instrument
only
shipping
1.72 kg (3.79 lb.)
5.22 kg (11.5 lb.), including standard accessories
Dimensions
45 mm x 204 mm x 331 mm (1.75 in x 8.0 in x 13 in) (H x W x D)
1
The humidity limits are derived from a psychrometric chart using a maximum wet bulb temperature of 29
_
C. The highest
RH (20%) at the maximum temperature (+60
_
C) diverges from the psychrometric chart (which would indicate 6%). The
20% number is a practical limit. Test chambers cannot be set below this number.
Summary of Contents for OI1125
Page 4: ......
Page 8: ...Table of Contents iv OI1125 E O Transmitter Instruction Manual...
Page 20: ...Getting Started 8 OI1125 E O Transmitter Instruction Manual...
Page 30: ...Operating Basics 18 OI1125 E O Transmitter Instruction Manual...
Page 36: ...Specifications 24 OI1125 E O Transmitter Instruction Manual...
Page 38: ......
Page 42: ...Theory of Operation 30 OI1125 E O Transmitter Instruction Manual...
Page 72: ...Performance Verification 60 OI1125 E O Transmitter Instruction Manual...