Performance Verification
52
OI1125 E/O Transmitter Instruction Manual
Oscilloscope Setup.
Set the TDS8000 Vertical settings as follows:
NOTE
. Due to differences in oscilloscope firmware, your setup steps may differ
slightly.
1. From the toolbar, select Setup
→
Vertical.
2. Click the Optical>> button.
3. Set the Signal Conditioning mode:
a. In the Wavelength field, select 1550 from the drop-down menu.
b. In the Filter field, select OC-48 from the drop-down menu.
NOTE
. You must change the Filter field setting when you take measurements at
other data rates. See Table 15.
A filter is not used for the 12.50 Gb/s data rate. Instead, set the Filter field to
None, and set the bandwidth to 12.50 GHz.
Table 15: Signal conditioning mode settings
Data rate,
Gb/s
Wavelength
field setting
Filter field
setting
Bandwidth
setting, GHz
2.488320
1550
OC-48
–––––––
9.953280
1550
OC-192
–––––––
12.500000
1550
None
12.50
4. Select the Trig tab to set the trigger parameters.
a. In the Trigger Source box, select External Direct.
Preview the Eye Pattern.
1. Turn on the BERT data and clock outputs.
2. Enable the tunable laser.
3. Enable the transmitter.
4. Turn on the channel that the optical module is plugged into.
5. Press Autoset.
The oscilloscope displays an eye pattern as shown in Figure 19 on page 54.
Summary of Contents for OI1125
Page 4: ......
Page 8: ...Table of Contents iv OI1125 E O Transmitter Instruction Manual...
Page 20: ...Getting Started 8 OI1125 E O Transmitter Instruction Manual...
Page 30: ...Operating Basics 18 OI1125 E O Transmitter Instruction Manual...
Page 36: ...Specifications 24 OI1125 E O Transmitter Instruction Manual...
Page 38: ......
Page 42: ...Theory of Operation 30 OI1125 E O Transmitter Instruction Manual...
Page 72: ...Performance Verification 60 OI1125 E O Transmitter Instruction Manual...