Performance Veri
fi
cation
Checking the Output Signal.
1.
Use the 75
Ω
BNC cable and 75
Ω
signal adapter to connect the ASI Output
connector on the MTX100B to the oscilloscope CH1 input.
Figure 2-19: Equipment connection for verifying the output signal-IEEE1394/ASI interface
2.
Set the oscilloscope controls as follows:
Control
Setting
Vertical scale
200 mV/div
Horizontal scale
1.25 ns/div
Acquire mode
Average 32
Trigger mode
AUTO
Trigger level
0 V
Trigger source
CH1
Trigger position
50%
Trigger slope
Rising Edge
Input impedance
50
Ω
Measure
Amplitude, Rise Time, Fall Time
Ref Level
High Ref 80%, Low Ref 20%
Gating
Cursor Curs1 Pos1: –2.5 ns/Curs2: 2.75 ns
MTX100B Speci
fi
cations and Performance Veri
fi
cation
2–37