Performance Veri
fi
cation
ASI Interface (Option 01
Only)
This test veri
fi
es that the transport stream data is correctly output from and is
recorded through the ASI interface on the MTX100B.
Required Equipment.
MPEG analyzer
Oscilloscope
75
Ω
signal adapter
test40.TRP
fi
le
Two 75
Ω
BNC cables
Checking the Output Signal.
1.
Use a 75
Ω
BNC cable and the 75
Ω
signal adapter to connect the ASI Out 1
connector on the MTX100B to the oscilloscope CH1 input.
Figure 2-11: Equipment connection for verifying the output signal-ASI interface
2.
Set the oscilloscope settings as follows:
Control
Setting
Vertical scale
200 mV/div
Input impedance
50
Ω
Horizontal scale
1.25 ns/div
Trigger position
50%
Acquire mode
Average 32
Trigger mode
AUTO
Trigger source
CH1