Performance Veri
fi
cation
Internal Clock Output Level
and Frequency Accuracy
This test veri
fi
es the internal clock output level and frequency accuracy.
Equipment required
Oscilloscope
Frequency counter
50
Ω
BNC cable
1.
Use the 50
Ω
BNC cable to connect the Trig In/Out connector on the
MTX100B to the oscilloscope CH1 input.
Figure 2-6: Equipment connection for verifying the internal clock output level
2.
Set the oscilloscope as indicated below:
Control
Setting
Vertical scale
1 V/div (CH1)
Input impedance
50
Ω
Horizontal scale
10 ns/div
Trigger position
50%
Acquire mode
Average 16
Trigger mode
Auto
Trigger level
1.20 V
Trigger source
CH1
Trigger slope
Rising Edge
Input coupling
DC
Measure
High Level, Low level
3.
Select
Play
>
Others
on the MTX100B to open the Others dialog box.
2–8
MTX100B Speci
fi
cations and Performance Veri
fi
cation
Summary of Contents for MTX100B
Page 2: ......
Page 6: ......
Page 12: ...General Safety Summary vi MTX100B Specifications and Performance Verification ...
Page 16: ...Introduction x MTX100B Specifications and Performance Verification ...
Page 17: ...Specifications ...
Page 18: ......
Page 36: ...Specifications 1 18 MTX100B Specifications and Performance Verification ...
Page 37: ...Performance Verification ...
Page 38: ......