Performance Veri
fi
cation
4.
Set the function generator settings as indicated below:
Control
Setting
Function
Square
Run Mode
Continuous
Frequency
10 MHz
Output Amplitude
0.5 V
Ch 1 output
Output Offset
2.0 V
Function
Arb (Edit: Point Number: 100, All Data:
16382)
Run Mode
Continuous
Output Menu
Load Impedance: 1 k
Ω
Output Amplitude
1.75 V
Ch 2 output
Output Offset
0.875 V
5.
Press the
On
button of the Ch1 output on the function generator.
6.
To open the
test64.TRP
fi
le on the MTX100B, do the following:
a.
Select
File
>
Open
in the Play screen to open the Select File dialog box.
b.
In the dialog box, select the
test64.TRP
fi
le.
7.
Select
Play
>
Update
>
On
.
8.
Select
Play
>
Clock
to open the Clock dialog box.
9.
In the dialog box, select
ExtRef 10
.
10.
Press the
Play/Pause
button on the MTX100B to start playing the
test64.TRP
fi
le.
11.
Verify that PLL unlock error does not occur on the MTX100B.
12.
Start the
TS Compliance Analyzer
on the MPEG analyzer.
13.
In the TS Compliance Analyzer window, select
Real-time..
, and then make
the following settings:
Control
Setting
Interfaces
DVB Parallel
Interface Settings
Time Stamping
14.
Verify that the hierarchical view is displayed on the MPEG analyzer screen
and that no error messages appear. In addition, verify that the data rate is
64
Mbps and that the packet size is
188
bytes.
NOTE.
Ignore Program 3 and PID 120 (0x78) errors in the Navigation view
because these are caused by the original test40.TRP
fi
le.
MTX100B Speci
fi
cations and Performance Veri
fi
cation
2–15
Summary of Contents for MTX100B
Page 2: ......
Page 6: ......
Page 12: ...General Safety Summary vi MTX100B Specifications and Performance Verification ...
Page 16: ...Introduction x MTX100B Specifications and Performance Verification ...
Page 17: ...Specifications ...
Page 18: ......
Page 36: ...Specifications 1 18 MTX100B Specifications and Performance Verification ...
Page 37: ...Performance Verification ...
Page 38: ......