![Tektronix KEITHLEY 4200A User Manual Download Page 68](http://html1.mh-extra.com/html/tektronix/keithley-4200a/keithley-4200a_user-manual_1078018068.webp)
Model 4200A-SCS Parameter Analyzer User's Manual
Section 5: Configure and use a Series 700 Switching System
4200A-900-01 Rev. C / January 2019
5-11
Set up the measurements in Clarius
After closing KCon, open the Clarius application from the desktop. In this section, you use the Clarius
application to configure and run two tests on an n-channel MOSFET transistor: A plot of drain current
versus drain voltage using the SMUs and a C-V sweep. By using the Series 700 Switching System,
you do not need to rearrange cables between the tests.
For this example, you use the Clarius application to:
Create a new project
Add a device
Add an action
Configure the action
Search for and add two tests
Run the project and view the tests
Create a new project
To create a new project:
1. Choose
Select
.
Figure 62: Select highlighted
2. In the Library, select
Projects
.
3. Select
New Project
.
4. Select
Create
.
Figure 63: Select a New Project from the Project Library