13. Select the arrow next to MTRX1 to see the settings for the 7072 Matrix Card.
14. Select
CARD1
.
15. Complete the Card Rows Assignments according to how you connected the instruments to the
7072. For this example, the assignments are:
Row A - SMU1 Force
Row B - SMU2 Force
Row C - SMU3 Force
Row E - GNDU Force
Row G - CVU1 CVH_CUR
Row H - CVU1 CVL_CUR
16. Under the Card Columns Assignment heading, designate at least the first four columns with pin
assignments that match their column number. For example, Pin 1 Force to column 1.
Figure 59: Completed Properties pane for the 7072 Matrix Card
17. From the KCon toolbar, select
Validate Configuration
to ensure that the switching system is
connected properly.
Figure 60: Validate Configuration icon