Section 8: Use the 4200A-CVIV Multi-Switch
Model 4200A-SCS Parameter Analyzer User's Manual
8-10
4200A-900-01 Rev. C / January 2019
Configure the action
To configure the action:
1. Select
Configure
.
Figure 112: Configure highlighted
2. In the project tree, select the
cviv-configure
action.
3. In the CVU 2/4 Wire Mode settings, set the output mode of the 4200A-CVIV to
Two Wire
.
4. In the CVIV Display Config settings, change the name of each channel to match the
corresponding MOSFET terminal. These names appear on the display of the 4200A-CVIV.
This step is not necessary for the test to run, but it allows you to see which terminal each output is
connected to on the 4200A-CVIV display.