Section 4: Source-measure hardware
Model 4200A-SCS Source-Measure Unit (SMU) User's Manual
4-10
4200A-SMU-900-01 Rev. A December 2020
FORCE terminal
The FORCE terminal is a standard triaxial connector used as a return path for the SMU or
preamplifier FORCE current. The center pin is FORCE, the inner shield is GUARD, and the outer
shield is circuit COMMON.
The ground unit FORCE and GUARD signal terminals are connected to circuit COMMON.
Figure 60: 4200-PA connectors