Model 4200A-SCS Source-Measure Unit (SMU) User's Manual
Section 3: Setting up SMUs in Clarius
4200A-SMU-900-01 Rev. A December 2020
3-33
Disable outputs at completion - SMU
If "Disable Outputs at Completion" is cleared, the SMU outputs remain at their last
programmed levels when the test is completed. To prevent electrical shock that could cause
injury or death, never make or break connections to the 4200A-SCS while the output is on.
When a test is complete, you can disable the outputs or leave them enabled. If you enable the
outputs, the SMU starts the next test from its state at the end of the previous test. If the next test does
not use the SMU, the SMU remains in that state.
To set the output action on completion:
1. Select the test.
2. Select
Configure
.
3. In the right pane, select the
Test Settings
tab.
4. Select
Advanced
.
5. Select or clear
Disable Outputs at Completion
.
Output Values
If you are using subsites, each time a subsite is cycled, the measurements for the enabled Output
Values are placed in the Analyze sheet for the subsite. For example, if the subsite is cycled five times,
there will be five measured readings (Output Values) for the test. For details, see
“Subsite cycling” in
the
Model 4200A-SCS Clarius User's Manual
.
To select output values:
1. In the project tree, select the test.
2. Select
Configure
.
3. In the right pane, select
Test Settings
.
4. Select
Output Values
. The Output Values dialog box is displayed.
5. Select the data series that you want to send to the Analyze sheet.
6. Select
OK
.