Appendix A: Remote programming
CH<n>:PRObe:FORCEDRange <NR3>
The command selects the dynamic range of the sensor tip in V
p-p
and is dependent
the attached sensor tip cable.
The following table lists the sensor tip cables and the dynamic range.
Table 14: Sensor tip cables and dynamic ranges
Sensor tip cable
Range V
p-p
1X
1.0 or 2.0
5X
5.0 or 10.0
10X
10.0 or 20.0
25X
25.0 or 50.0
50X
50.0 or 100.0
CH<n>:PRObe:FORCEDRange? The query returns the dynamic range of the
sensor tip in V
p-p
.
CH<n>:PRObe:GAIN?
Returns the gain factor of the sensor tip (inverse of attenuation). Dependent on the
attached sensor tip cable. Query only.
CH<n>:PRObe:ID {:SERnumber | :TYPe}?
Query only. When sending only PRObe:ID?, the query returns the sensor tip type
string followed by the serial number string.
When sending PRObe:SERnumber?, the query returns the serial number string.
When sending PRObe:TYPe?, the query returns one of the following sensor tip
type strings (note the trailing spaces within the return strings):
“TIVM1
”(1GHz, 3m)
“TIVM1L ”(1GHz, 10m)
“TIVM05 ”(500MHz, 3m)
“TIVM05L”(500MHz, 10m)
“TIVM02 ”(200MHz, 3m)
“TIVM02L”(200MHz, 10m)
56
TIVM Series Measurement System User Manual
Summary of Contents for IsoVu TIVM Series
Page 1: ...xx TIVM Series IsoVu Measurement System ZZZ User Manual P071349500 071 3495 00...
Page 2: ......
Page 6: ......
Page 18: ...Preface xii TIVM Series Measurement System User Manual...
Page 40: ...Operating information 22 TIVM Series Measurement System User Manual...
Page 46: ...Application examples 28 TIVM Series Measurement System User Manual...
Page 72: ...User service 54 TIVM Series Measurement System User Manual...
Page 80: ...Appendix B Compliance information 62 TIVM Series Measurement System User Manual...