User service
Clean the exterior surfaces with a dry, lint-free cloth or a soft-bristle brush. If
dirt remains, use a soft cloth or swab dampened with a 75% isopropyl alcohol
solution. Use only enough solution to dampen the cloth or swab. Do not use
abrasive compounds on any part of the instrument.
Performance veri
fi
cation procedures
Use the following procedures to verify the performance of the IsoVu measurement
system. Before beginning the procedures, photocopy the test record and use it to
record the performance results. (See page 53,
Required equipment
The equipment required to perform the performance veri
fi
cation procedures are
shown in the following table.
Table 11: Required equipment for performance veri
fi
cation
Description
Minimum requirements
Example product
Oscilloscope with TekVPI interface
≥
1 GHz, 50
Ω
input support, fully compatible
with TekVPI interface
Tektronix MDO4104C
Pulse generator
1 V
p-p
, <1ns risetime
Tektronix Tek-DPG
TIVM series 1X sensor tip cable
As per description
Tektronix IVTIP1X
MMCX Jack (female) to BNC
female adapter
As per description
Fairview Microwave Product SKU: SM3610
Preparation
Prepare the equipment as follows:
1.
Turn on the TekVPI oscilloscope.
2.
Connect the Tek-DPG Deskew pulse generator into Channel 2 of the TekVPI
oscilloscope.
3.
Allow the test equipment to warm up for 20 minutes at an ambient temperature
of 20 °C to 30 °C.
46
TIVM Series Measurement System User Manual
Summary of Contents for IsoVu TIVM Series
Page 1: ...xx TIVM Series IsoVu Measurement System ZZZ User Manual P071349500 071 3495 00...
Page 2: ......
Page 6: ......
Page 18: ...Preface xii TIVM Series Measurement System User Manual...
Page 40: ...Operating information 22 TIVM Series Measurement System User Manual...
Page 46: ...Application examples 28 TIVM Series Measurement System User Manual...
Page 72: ...User service 54 TIVM Series Measurement System User Manual...
Page 80: ...Appendix B Compliance information 62 TIVM Series Measurement System User Manual...