![Tektronix IsoVu TIVM Series User Manual Download Page 29](http://html1.mh-extra.com/html/tektronix/isovu-tivm-series/isovu-tivm-series_user-manual_1078002029.webp)
Operating information
The following steps describe the process for connecting the measurement system
between a Tektronix oscilloscope and the DUT.
1.
Verify the DUT is not connected to an energized circuit.
2.
Connect the Comp box to one of the channels on the oscilloscope.
Figure 6: Connect the Comp box to the oscilloscope
3.
Refer to the following
fi
gure and install the sensor tip cable to the sensor head.
a.
Line up the sensor tip cable with the sensor head.
Take care to avoid bending or twisting the sensor tip cable assembly
during this process.
b.
Connect the SMA connector of the sensor tip cable to the sensor head.
Use the accessories wrench to tighten the SMA cable to 4 to 5-in lbs.
Use the adjustment tool that was shipped with your system.
c.
Press the sensor tip cable housing against the sensor head and then tighten
the two screws to 3 to 5-in lbs.
Use the adjustment tool that was shipped with your system.
Figure 7: Connecting the sensor tip cable to the sensor head
TIVM Series Measurement System User Manual
11
Summary of Contents for IsoVu TIVM Series
Page 1: ...xx TIVM Series IsoVu Measurement System ZZZ User Manual P071349500 071 3495 00...
Page 2: ......
Page 6: ......
Page 18: ...Preface xii TIVM Series Measurement System User Manual...
Page 40: ...Operating information 22 TIVM Series Measurement System User Manual...
Page 46: ...Application examples 28 TIVM Series Measurement System User Manual...
Page 72: ...User service 54 TIVM Series Measurement System User Manual...
Page 80: ...Appendix B Compliance information 62 TIVM Series Measurement System User Manual...