Operating information
Table 4: Environmental considerations (cont.)
Feature
Description
Sensor tip cables/adapters
Operating
5% to 80% RH (Relative Humidity), at up to 40 °C (104 °F)
5% to 45 % RH above 40 °C (104 °F) up to 85°C (185 °F),
non-condensing
Non operating
5% to 85% RH (Relative Humidity), at up to 40 °C (104 °F)
5% to 45 % RH above 40 °C (104 °F) up to 70 °C (158 °F),
non-condensing
Altitude
Operating
3000 m (9843 ft.)
Non operating
12,000 m (39370 ft.)
Clearance requirements
The unique common mode voltage range of the measurement system allows it to
be used in the presence of high frequency/high voltage common mode signals. It
is important to observe all precautions while using this product.
WARNING.
RF burns can occur while using this measurement system. Anyone
making measurements that might come within the RF burn regions indicated in
the following
fi
gure should be familiar with the hazards of working with signals in
these regions and take appropriate action, such as RF shielding for the DUT.
4
TIVM Series Measurement System User Manual