Operating information
Environmental
requirements
The following tables describe speci
fi
cations and maximum operating
environmental ratings for the measurement system when connected to a DUT
and a Tektronix oscilloscope.
Table 3: Input speci
fi
cations
Feature
Description
1000 V
rms
, CAT II
Common mode
2000 V peak, CAT I (expected transient voltage 2500 V peak)
Differential mode
Sensor tip cable dependent (See the Sensor tip cable voltage
rating below.)
Table 4: Environmental considerations
Feature
Description
Temperature
Controller
Operating
0 °C to 40 °C (32 °F to 104 °F)
Non-operating
-40 °C to 70 °C (-40 °F to 158 °F)
Sensor head
Operating
0 °C to 70 °C (32 °F to 158 °F)
Non-operating
-40 °C to 70 °C (-40 °F to 158 °F)
Sensor tip cables/adapters
Operating and
non-operating
-40 °C to 85 °C (-40 °F to 185 °F)
Humidity
Controller
Operating
5% to 85% RH (Relative Humidity), at up to 40 °C (104 °F),
non-condensing
Non operating
5% to 85% RH (Relative Humidity), at up to 40 °C (104 °F)
5% to 45 % RH above 40 °C (104 °F) up to 70 °C (158 °F),
non-condensing
Sensor head
Operating
5% to 80% RH (Relative Humidity), at up to 40 °C (104 °F)
5% to 45 % RH above 40 °C (104 °F) up to 70 °C (158 °F),
non-condensing
Non operating
5% to 85% RH (Relative Humidity), at up to 40 °C (104 °F)
5% to 45 % RH above 40 °C (104 °F) up to 70 °C (158 °F),
non-condensing
TIVM Series Measurement System User Manual
3
Summary of Contents for IsoVu TIVM Series
Page 1: ...xx TIVM Series IsoVu Measurement System ZZZ User Manual P071349500 071 3495 00...
Page 2: ......
Page 6: ......
Page 18: ...Preface xii TIVM Series Measurement System User Manual...
Page 40: ...Operating information 22 TIVM Series Measurement System User Manual...
Page 46: ...Application examples 28 TIVM Series Measurement System User Manual...
Page 72: ...User service 54 TIVM Series Measurement System User Manual...
Page 80: ...Appendix B Compliance information 62 TIVM Series Measurement System User Manual...