Commands Listed in Alphabetical Order
MINImum
fi
nds the minimum amplitude. This value is typically the most negative
peak voltage. It is measured over the entire waveform or gated region. This
measurement is applicable only to the analog channels.
NDUty
(negative duty cycle) is the ratio of the negative pulse width to the signal
period, expressed as a percentage. The duty cycle is measured on the
fi
rst cycle in
the waveform or gated region.
Negative Duty Cycle = ((Negative Width) / Period) × 100%
NEDGECount
is the count of negative edges.
NOVershoot
(negative overshoot)
fi
nds the negative overshoot value over the
entire waveform or gated region. This measurement is applicable only to the
analog channels.
Negative Overshoot = ((Low - Minimum) / Amplitude) × 100%)
NPULSECount
is the count of negative pulses.
NWIdth
(negative width) measurement is the distance (time) between the middle
reference (default = 50%) amplitude points of a negative pulse. The measurement
is made on the
fi
rst pulse in the waveform or gated region.
PEAKHits
measures the number of points in the largest bin of the histogram.
PDUty
(positive duty cycle) is the ratio of the positive pulse width to the signal
period, expressed as a percentage. It is measured on the
fi
rst cycle in the
waveform or gated region.
Positive Duty Cycle = ((Positive Width)/Period) × 100%
PEDGECount
is the count of positive edges.
PERIod
is the time required to complete the
fi
rst cycle in a waveform or gated
region. Period is the reciprocal of frequency and is measured in seconds.
PHAse
measures the phase difference (amount of time a waveform leads or lags
the reference waveform) between two waveforms. The measurement is made
between the middle reference points of the two waveforms and is expressed in
degrees, where 360° represents one waveform cycle.
PK2Pk
(peak-to-peak)
fi
nds the absolute difference between the maximum and
minimum amplitude in the entire waveform or gated region. This measurement is
applicable only to the analog channels.
POVershoot
is the positive overshoot value over the entire waveform or gated
region. This measurement is applicable only to the analog channels.
Positive Overshoot = ((Maximum - High) / Amplitude) ×100%
PPULSECount
is the count of positive pulses.
MSO4000 and DPO4000 Series Programmer Manual
2-235
Summary of Contents for DPO4000 Series
Page 2: ......
Page 10: ...Table of Contents vi MSO4000 and DPO4000 Series Programmer Manual ...
Page 23: ...Command Syntax MSO4000 and DPO4000 Series Programmer Manual 2 9 ...
Page 24: ...Command Syntax 2 10 MSO4000 and DPO4000 Series Programmer Manual ...
Page 84: ...Command Groups 2 70 MSO4000 and DPO4000 Series Programmer Manual ...
Page 546: ...Commands Listed in Alphabetical Order 2 532 MSO4000 and DPO4000 Series Programmer Manual ...
Page 564: ...Status and Events 3 18 MSO4000 and DPO4000 Series Programmer Manual ...
Page 565: ...Appendix A Character Set MSO4000 and DPO4000 Series Programmer Manual A 1 ...
Page 566: ...Appendix A Character Set A 2 MSO4000 and DPO4000 Series Programmer Manual ...
Page 572: ...Appendix B Reserved Words B 6 MSO4000 and DPO4000 Series Programmer Manual ...