b. CHECK-the displayed GATE light blinks and the
display readout is 0.0 (SEC)
count with the display
Hz/SEC illuminated.
11. Check RATIO A/B Minimum Pulse Width (4 ns
at 100 mV peak-to-peak).
a. Set the DC 503A FUNCTION switch to RATIO A/B.
b. CHECK-the displayed GATE light blinks and the
display readout is 0.000000
count (no annunciator
lights).
12. Check the Input Sensitivity X1 Attenuation
(60 mV p-to-p pulse at minimum pulse width of 5 ns
to 100 MHz). Refer
control settings as shown in step 9.
a. Remove the DC 503A B SHAPED OUT connection
from the vertical plug-in INPUT.
b. Change the coaxial cable (with the 10X attenuator)
from the DC 503A CH B INPUT to the vertical plug-in
INPUT.
c. Change the vertical plug-in VOLTS/DIV to 20 mV.
Adjust the pulse generator OUTPUT (VOLTS) Low
d.
LEVEL control to position the bottom edge
of
the dis-
played Squarewave to the center of the CM graticule.
e. Adjust the pulse generator OUTPUT (VOLTS) HIGH
EDGE control for three divisions of display (60 mv p-to-p)
on the crt.
f. Change the vertical plug-in VOLTS/DIV to 0.1 and
adjust the pulse generator PERIOD VARIABLE for a
period of 10 ns (5 divisions).
g. Move the vertical plug-in INPUT connection to the
DC 503A CH A INPUT and connect the A SHAPED OUT
signal to the vertical plug-in INPUT.
h. Adjust the DC 503A CH A LEVEL control for a stable
display
on
the DC 503A and oscilloscope.
i. CHECK-that the DC 503A readout indicates ap-
proximately 100.00 (MHz) with the display
TM 9-6625-474-14&P-3
j. Set the DC 503A FUNCTION switch to PERIOD B
(AVGS) and the AVGS switch to
k. Move the DC 503A A SHAPED OUT connector to
the B SHAPED OUT (black terminal to COMMON) and
move the CH A INPUT connection to the CH B lNPUT.
l. Adjust the DC 503A CH B LEVEL control for a stable
display on the DC 503A and oscilloscope.
m. CHECK-that the DC 503A readout indicates ap-
proximately 10.000 (nSEC) with the display GHz/nSEC
illuminated.
13. Check the Width B (AVGS) Minimum Pulse
Width (5 ns).
a. Set the DC 503A FUNCTION switch to WIDTH B
(AVGS).
b. CHECK-the displayed GATE light blinks and the
readout indicates approximately 5.0000 (nSEC) with the
display GHz/nSEC illuminated.
14. Check the Events A During B Minimum B Pulse
Width (5 ns).
a. Set the DC 503A FUNCTION switch to EVENTS A
DUR B.
b. CHECK-the displayed GATE light blinks and the
display readout is 0.000000
count (no annunciator
lights).
15. Check the Width B (no AVGS) Minimum Pulse
Width (20 ns).
a. Change the DC 503A FUNCTION switch to PERIOD
B (AVGS).
b. Change the pulse generator PERIOD to 10 ns and
adjust PERIOD VARIABLE for a DC 503A display readout
of approximately 40.0000 (nSEC) with the display
GHz/nSEC illuminated.
c. Change DC 503A FUNCTION switch to PERIOD B
(no AVGS).
illuminated.
Summary of Contents for DC503a
Page 1: ...e x _ 1 v w W i L ...
Page 16: ...DC 503A DC 503A Universal Counter Timer REV A FEB 1981 XIV ...
Page 27: ...Fig 2 2 Controls and connectors Operating Instructions DC 503A 2977 fli English 2 3 ...
Page 39: ...Fig 2 2 Commandes et prises Instructions d utilisation DC 503A French 2 3 ...
Page 51: ...Abb 2 2 Bedienungselemente and Anschliisse Bedienungsanleitung DC 503A German 2 3 ...
Page 63: ...Fig 2 2 Controls and connectors Operating Instructions DC 503A 2971 01 Japanese 2 3 ...
Page 125: ...Fig 8 2 Auxiliary Board A12 Assy ADJUSTMENT LOCA 2971 1 6 ...
Page 169: ...SEE END OF MPL FOR WIRE ASSEMBLIES O1 0 pG C 0 DC 503A ...
Page 186: ...0 2 TM 9 6625 474 14 P 3 Fig 0 1 DC 503A Universal Counter Timer ...
Page 188: ...Table 1 1 1 2 TM 9 6625 474 14 P 3 ...
Page 189: ...Table 1 1 1 3 TM 9 6625 474 14 P 3 ...
Page 190: ...Table 1 1 1 4 TM 9 6625 474 14 P 3 ...
Page 191: ...Table 1 1 1 5 TM 9 6625 474 14 P 3 ...
Page 192: ...1 6 Table 1 1 TM 9 6625 474 14 P 3 ...
Page 193: ...Table 1 1 1 7 Table 1 2 Table 1 3 TM 9 6625 474 14 P 3 ...
Page 194: ...Table 1 3 Table 1 4 1 8 TM 9 6625 474 14 P 3 ...
Page 197: ...2 3 TM 9 6625 474 14 P 3 Fig 2 2 Controls and connectors ...
Page 211: ...3 5 Fig 3 1 TM 9 6625 474 14 P 3 ...
Page 219: ...4 2 TM 9 6625 474 14 P 3 Table 4 1 LIST OF TEST EQUIPMENT REQUIREMENTS ...
Page 240: ...5 7 TM 9 6625 474 14 P 3 Fig 5 9 Rear interface connector assignments ...
Page 256: ...8 3 8 4 blank TM 9 6625 474 14 P 3 ...
Page 257: ...8 5 8 6 blank TM 9 6625 474 14 P 3 ...
Page 258: ...Fig 8 2 Fig 8 1 8 7 8 8 blank TM 9 6625 474 14 P 3 ADJUSTMENT LOCATIONS ...
Page 260: ...TM 9 6625 474 14 P 3 8 11 8 12 blank ...
Page 262: ...TM 9 6625 474 14 P 3 8 15 8 16 blank ...
Page 263: ...8 17 8 18 blank TM 9 6625 474 14 P 3 Table 8 3 COMPONENT REFERENCE CHART See Fig 8 3 ...
Page 264: ...8 19 8 20 blank TM 9 6625 474 14 P 3 ...
Page 265: ...8 21 8 22 blank TM 9 6625 474 14 P 3 Table 8 4 COMPONENT REFERENCE CHART See Fig 8 3 ...
Page 266: ...8 23 8 24 blank TM 9 6625 474 14 P 3 ...
Page 267: ...8 25 8 26 blank TM 9 6625 474 14 P 3 DC 503A Table 8 5 COMPONENT REFERENCE CHART See Fig 8 4 ...
Page 268: ...8 27 8 28 blank TM 9 6625 474 14 P 3 ...
Page 269: ...8 29 8 30 blank TM 9 6625 474 14 P 3 DC 503A Table 8 6 COMPONENT REFERENCE CHART See Fig 8 3 ...
Page 270: ...8 31 8 32 blank TM 9 6625 474 14 P 3 ...
Page 271: ...8 33 8 34 blank TM 9 6625 474 14 P 3 DC 503A Table 8 7 COMPONENT REFERENCE CHART See Fig 8 4 ...
Page 272: ...8 35 8 36 blank TM 9 6625 474 14 P 3 ...
Page 274: ...8 39 8 40 blank TM 9 6625 474 14 P 3 ...
Page 275: ...8 41 8 42 blank TM 9 6625 474 14 P 3 DC 503A Table 8 9 COMPONENT REFERENCE CHART See Fig 8 3 ...
Page 276: ...8 43 8 44 blank TM 9 6625 474 14 P 3 ...
Page 277: ...8 45 8 46 blank TM 9 6625 474 14 P 3 DC 503A Table 8 10 COMPONENT REFERENCE CHART See Fig 8 4 ...
Page 278: ...8 47 8 48 blank TM 9 6625 474 14 P 3 ...
Page 291: ...D 1 TM 9 6625 474 14 P 3 APPENDIX D MANUAL CHANGE INFORMATION ...
Page 292: ...D 2 TM 9 6625 474 14 P 3 ...
Page 293: ...D 3 TM 9 6625 474 14 P 3 P a g e 1 o f 2 ...
Page 294: ...Page 2 D 4 TM 9 6625 474 14 P 3 ...
Page 296: ...D 6 TM 9 6625 474 14 P 3 ...
Page 298: ......
Page 299: ...THE METRIC SYSTEM AND EQUIVALENTS ...
Page 300: ...PIN 056816 000 ...