Appendix F: User Service
F–18
DAS System User Manual
Red-Striped Wire
Pin 1
Figure F–5: Installing the GPIB Connector
Replacing the Line Fuse
The following procedure describes how to change the line fuse for the main-
frame. To change the fuse, perform the following steps:
1. Power off the system unit.
2. Wait 60 seconds and disconnect the power cord from the system unit. This
allows time for the power-down sequence to complete.
3. Remove the line fuse and replace it with the appropriate fuse. Table F–3 lists
the fuses and the Tektronix part number.
Table F–3: System Unit Fuse Replacement
Line Operation
Fuse
Tektronix Part Number
115 VAC
15 A, 250 V, Fast blow
159-0256-00
230 VAC
10 A, 250 V, 3AG
159-0057-00