Appendix C: Error Messages
C–2
DAS System User Manual
Table C–1: Error Messages (Cont.)
Error Message
Explanation
Can’t find selected symbol table file.
Symbol table file does not exist on the hard disk; verify you are using correct file name
or install the symbol table file.
Can’t save to Tek-defined Library Items.
You cannot save to any of the predefined Library Items; you can only save to Library
Items that you have defined.
Cannot delete first line of clock equation.
You cannot delete the first line; you may add or delete a second line.
Cannot deskew if the module is running or
hardware is otherwise busy.
Deskew requires that the module not be busy with other activities. Wait until the
module is available.
Cannot use external components with
High-Speed signals.
You cannot use high-speed signals for external input or output. Only normal signals
can be used.
Caution: Signal [signal name]:High Speed
Signal line cannot be allocated.
Line cannot be allocated; an empty slot exists between the modules or they do not
support high-speed signals.
Channel [channel number] already being used
in the group [group name].
The selected channel number is already being used in another group.
Channel names must be unique.
No two channel names can be the same. The old channel name will be restored.
Characters-per-line value too large for
PostScript, set to maximum.
PostScript allows a maximum of 80 characters-per-line.
Choose another name; Default is reserved.
Enter a name other than Default.
Clock edge is never active.
The 92A96 clock qualifiers have been logically ANDed together such that the clock
edge is never active.
Clock edge is always active.
The 92A96 clock qualifiers have been logically ORed together such that the clock
edge is always active.
Cluster: [cluster name] will be destroyed.
Press Return to confirm, Esc to abort.
You have assigned a module to more than one cluster. If you press the Return key,
the specified cluster will be destroyed.
Clusters must contain at least one module. No
changes will be saved. Press Return to
confirm, Esc to abort.
You’ve defined a cluster with no modules; this cluster definition cannot be saved. If
you press the Return key, the cluster definition will be deleted.
Compound Event cannot contain 2 Counters
or Timers.
A Compound Event can contain only one counter or timer.
Conflict with Event in last State.
The “Store Event” and “Combination” selections in the storage field are not valid if the
third clause in the last (fourth) State is currently in use.
Conversion destination is not wide enough.
The width of the conversion destination must be equal to or greater than the source
width.
Counter not available, 2 Timers in use.
Only two counter/timer resources are available.
Counter or Timer unavailable, 2 Counters
and/or Timers maximum.
Only two counter/timer resources are available.
Current clock equation fully expanded.
You can have a maximum of four lines in a 92A96 clock equation.
Cursor must be in an Action Macro.
Position cursor on the Action Macro you want to change and press the function key
again.
Cursor must be in an Event Macro.
Position cursor on the Event Macro you want to change and press the function key
again.
Summary of Contents for DAS 9221 Option 04
Page 4: ......
Page 6: ......
Page 16: ...Table of Contents x DAS System User Manual ...
Page 20: ...General Safety Summary xiv DAS System User Manual ...
Page 23: ......
Page 48: ......
Page 81: ......
Page 156: ......
Page 166: ...Appendix B Accessories B 4 DAS System User Manual ...
Page 178: ...Appendix C Error Messages C 12 DAS System User Manual ...
Page 238: ...Appendix F User Service F 26 DAS System User Manual ...
Page 249: ......
Page 266: ......
Page 275: ......
Page 276: ......