Command Descriptions
2-98
AWG710 Arbitrary Waveform Generator Programmer Manual
TRIGger:SEQuence:LEVel 200mV
sets the trigger level to 200 mV.
TRIGger[:SEQuence]:POLarity (?)
This command selects the polarity relative to the trigger level that is required to
activate the gate signal. This command is effective only when the waveform
generator is in the gated mode.
Trigger
AWGControl:RMODe, TRIGger[:SEQuence]:LEVel
TRIGger[:SEQuence]:POLarity { Positive | Negative }
TRIGger:POLarity?
:
TRIGger
SEQuence
:
POLarity
<space>
?
Positive
Negative
Positive
means the gate signal is activated when the external trigger signal is
greater (more Positive) than the trigger level.
Negative
means the gate signal is activated when the external trigger signal is
less (more Negative) than the trigger level.
At
*RST
, the parameter is set to Positive.
TRIGger[:SEQuence]:POLarity Negative
selects the
Negative
polarity.
Examples
Group
Related Commands
Syntax
Arguments
Examples
Summary of Contents for AWG710
Page 4: ......
Page 14: ...Preface x AWG710 Arbitrary Waveform Generator Programmer Manual...
Page 15: ...Getting Started...
Page 16: ......
Page 28: ...Getting Started 1 12 AWG710 Arbitrary Waveform Generator Programmer Manual...
Page 29: ...Syntax and Commands...
Page 30: ......
Page 134: ...Retrieving Response Messages 2 104 AWG710 Arbitrary Waveform Generator Programmer Manual...
Page 145: ...Status and Events...
Page 146: ......
Page 173: ...Examples...
Page 174: ......
Page 176: ...Programming Examples 4 2 AWG710 Arbitrary Waveform Generator Programmer Manual...
Page 177: ...Appendices...
Page 178: ......
Page 195: ...Glossary and Index...
Page 196: ......
Page 198: ...Glossary Glossary 2 AWG710 Arbitrary Waveform Generator Programmer Manual...
Page 204: ...Index Index 6 AWG710 Arbitrary Waveform Generator Programmer Manual...
Page 205: ......
Page 206: ......