VOLTAGE
AND WAVEFORM
CONDITIONS
WARNING
Dangerous potentials exist
at
several points throughout this instrument. When the
instrument
is
operated
with the covers
removed, do not touch exposed connections or
components.
Some transistors have voltages present on their cases. Disconnect the
power source before
replacing parts.
RECOMMENDED TEST EQUIPMENT
Item
Specifications
Recommended
Type
Test oscilloscope system
Deflection factor,
1
mV to 50 V/div; input
impedance, 1
megohm; frequency
response,
de
to 2 MHz.
Probe: 10X
attenuation
probe
compatible with
vertical
input.
Tektronix
5110, 5A13N, 5B10N
oscilloscope system or
equiv.
Use a
Tektronix
P6108 or P6062B Probe.
Voltmeter (Non-loading
digital multimeter)
Range,
0 to 250 V; input impedance,
10
megohms.
Tektronix DM
501A
Option
02 Digital
Multimeter with power module.
VOLTAGE
CONDITIONS
Voltage
measurements
on this diagram were
made under the following conditions:
An
amplifier unit is
installed
in the left vertical compartment (for power supply loading). INTENSITY control is set fully
ccw.
Voltmeter common is
connected to chassis ground.
WAVEFORM
CONDITIONS
OSCILLOSCOPE UNDER TEST.
Install
an amplifier unit in the left vertical compartment and a time-base unit in the
horizontal compartment.
Connect
the CALIBRATOR output signal to the amplifier unit (set vertical input coupling to de
and
volts/div
for
a 2-division display). Set the time-base unit for internal auto-trigger,
2
ms/division sweep rate.
TEST
OSCILLOSCOPE.
Set
the test
oscilloscope triggering for auto mode with ac coupling from the internal
source and
set vertical
input coupling to ac. Connect a 10X Probe to the vertical
input. Position the display as necessary.
NOTE
The
waveforms shown are actual waveform photographs
taken with a Tektronix Oscilloscope Camera System and
Projected Graticule. Vertical
deflection factor shown on the waveform is the actual deflection factor from the probe tip.
Voltages and
waveforms on
the diagrams are not absolute and may vary between instruments because of component
tolerances,
internal calibration, or front-panel settings. Readouts are simulated in
iarger-than-normal type.
0
©
<D
Summary of Contents for 5110
Page 6: ...5110 Fig 1 1 5110 Oscilloscope 2134 01 ...
Page 22: ......
Page 34: ......
Page 42: ......
Page 83: ...5110 I43XII3 2134 15 ...
Page 84: ... i 143 I132 2134 15 BLOCK DIAGRAM ...
Page 85: ... t 2 E I 5110 S LNHNOdWOD BOVJHaiNI ...
Page 88: ...J4 O3 INTERFACE J INTERFACE A NOV 1978 ...
Page 92: ...AUXILIARY BOARD 1143 2134 18 AUXILIARY BOARD ...
Page 95: ...DEFLECTION AMPLIFIERs ...
Page 102: ...LV POWERSUPPLY CALIB COMPONENTS A 4 L V Power Supply circuit board 5110 ...
Page 107: ...SIGNALS OUT OPTION 7 ONLY SI 10 OPT 7 213 26 REV A FE BI979 SIGNALS OUT COPTION T ONLY n ...
Page 114: ......
Page 115: ...5110 OSCILLOSCOPE ...
Page 116: ...FIG 2 MAINFRAME p 5110 OSCILLOSCOPE ...
Page 117: ...1 ...
Page 121: ...o ...
Page 122: ...S 1 5110 OSCILLOSCOPE ...
Page 123: ...13 FIG 4 RACKMOUNT CABINET 5110 OSCILLOSCOPE ...
Page 124: ......