Section
1—5110
OPERATING
INSTRUCTIONS
This
instruction manual provides both operating and
servicing
information
for
the 5110 Oscilloscope. The
manual is divided into nine
sections. Operating, specifica
tion, and
performance
check information is covered in the
first
two sections, and is intended for operating and
service
personnel. Servicing information is covered in the
remaining
sections of the manual, and is intended for
qualified service personnel only.
OPERATING
POWER
This
instrument can be
operated from eithera 120-volt
or240-volt nominal line-voltage source,
48 to 440 hertz. In
addition, three
regulating
ranges are
provided for each
nominal
line-voltage
source.
PRELIMINARY
INFORMATION
Oscilloscope
Features
The
oscilloscope is a
solid
state,
light weight instru
ment
designed
for general-purpose measuring
applications. This
instrument has three plug-in com
partments
that accept plug-in
units to form
a complete
measurement
system. The
two-plug-in
compartments on
the
left are
connected to the vertical deflection system.
The
right
plug-in compartment is connected
to the
horizontal
deflection
system.
Electronic switching
between
the
vertical plug-in compartments allows a multi
trace
vertical
display. The flexibility of this plug-in feature
and
the variety of plug-in units available allow this system
to
be used for many measurement applications.
This
instrument
features a large-screen, 8X10 division
display; each
division equals 0.5 inch
(1.27 centimeter).
Regulated d
c
power supplies ensure that performance is
not affected
by variations
in line voltage and
frequency, or
by changes
in
the load due to
the varying power re
quirements
of the plug-in units.
Safety
Information
This instruction
manual contains warning information
which
the user
must follow
to
ensure safe operation of the
instrument.
Warning
information is
intended to protect the
operator
and Caution
information is intended to protect
the
instrument.
To
prevent
damage to the instrument, always check
the
line-voltage information recorded on the rear
panel
before applying power to the instrument.
WARNIWG~|
This
instrument
is intended to
be operated from
a
single-phase
earth-referenced power source having
one
current-carrying conductor (the
Neutral Con
ductor)
near
earth potential.
Operation from power
sources
where
both current-carrying conductors
are
live with respect to earth (such as phase-to-
phase on a three-wire system) is not recommended,
since
only the Line Conductor has over-current
(fuse)
protection within the instrument.
This
instrument
has a three-wire power cord with a
polarized
two-pole,
three-terminal plug for connec
tion
to the power source and safety-earth. The
safety-earth
terminal of
the plug is
directly con
nected
to
the instrument frame. For electric-shock
protection, insert this plug
only in a mating outlet
with
a
safety-earth contact.
Do not defeat
the
grounding connection. Any in
terruption
of the grounding connection can create
an
electric-shock hazard.
Before
making external
connections to
this instrument, always ground the
instrument
first by
connecting the power-cord to
a
proper
mating power outlet.
WARNING
I
High
voltage is present
inside
the instrument. To
avoid
electric-shock
hazard,
operating personnel
must not
remove the protective instrument covers.
Component replacement and
internal
adjustments
must be
made
by qualified service personnel only.
OPERATING
TEMPERATURE
The
instrument can be operated
where the ambient air
temperature
is
between 0Q C and
+50° C. The instrument
can
be stored in ambient
temperature between — 40° C and
+70°
C.
After
storage at a
temperature beyond the
operating
limits,
allow the chassis temperature to come
within
the operating
limits before power is
applied.
1-1
Summary of Contents for 5110
Page 6: ...5110 Fig 1 1 5110 Oscilloscope 2134 01 ...
Page 22: ......
Page 34: ......
Page 42: ......
Page 83: ...5110 I43XII3 2134 15 ...
Page 84: ... i 143 I132 2134 15 BLOCK DIAGRAM ...
Page 85: ... t 2 E I 5110 S LNHNOdWOD BOVJHaiNI ...
Page 88: ...J4 O3 INTERFACE J INTERFACE A NOV 1978 ...
Page 92: ...AUXILIARY BOARD 1143 2134 18 AUXILIARY BOARD ...
Page 95: ...DEFLECTION AMPLIFIERs ...
Page 102: ...LV POWERSUPPLY CALIB COMPONENTS A 4 L V Power Supply circuit board 5110 ...
Page 107: ...SIGNALS OUT OPTION 7 ONLY SI 10 OPT 7 213 26 REV A FE BI979 SIGNALS OUT COPTION T ONLY n ...
Page 114: ......
Page 115: ...5110 OSCILLOSCOPE ...
Page 116: ...FIG 2 MAINFRAME p 5110 OSCILLOSCOPE ...
Page 117: ...1 ...
Page 121: ...o ...
Page 122: ...S 1 5110 OSCILLOSCOPE ...
Page 123: ...13 FIG 4 RACKMOUNT CABINET 5110 OSCILLOSCOPE ...
Page 124: ......