Adjustment-5110
TIME-BASE PLUG-IN
Display
Position
Seconds/Div
Seconds/Div
Cal
Swp Mag
Triggering
Triggering Source
Chop
Centered
1
ms
Fully
clockwise
Off
+ Slope, Auto Trig,
ac
Coupl
Composite
ADJUSTMENT
PROCEDURE
1. Adjust
—30 Volt
Power Supply
a. Turn over the oscilloscope to
lay on its left side to
gain
access to the LV Power Supply circuit board.
b.
Connect
the digital
voltmeter between the —30 V
test point
and ground. See
Fig. 3-1 for voltage test point
location.
c.
Check—
for a
meter reading of —29.89 to —30.11
volts.
NOTE
If
the 30
volt supply is within the specified
tolerance,
proceed with step 2. If
the —30 volt
adjustment
is to be made, all circuits will be affected
and
the entire
power supply adjustment procedure
should be performed to
verify the accuracy of the
supplies.
d.
Adjust-----
30 V Adj
R878 for a meter
reading of
exactly —30 volts.
See Fig. 3-1 for adjustment location.
2. Adjust
+30 Volt Power Supply
a. Connect the digital voltmeter
between the +30 V test
point and ground.
See Fig.
3-1 for
voltage test point
location.
b.
Check—
for a
meter
reading of +29.82 to +30.18
volts.
NOTE
If the
+30 volt
supply
is within the specified
tolerance,
proceed with step
3. If the
+30
volts
adjustment is
to
be
made, all circuits will be affected
and
the
entire power supply
adjustment procedure
should
be performed to verify
the accuracy of the
supplies.
c.
30 V Adj
R858 for
a meter
reading of
e30 volts. See Fig. 3-1 for adjustment location.
3.
Check Remaining
Power
Supply Voltages
a. Connect the digital voltmeter between the +5 V test
point and ground.
See Fig. 3-1 for voltage
test point
location.
b.
Check—
for a meter
reading of +4.89 to +5.11 volts.
c.
Connect the
digital voltmeter between the +200 V
test point
and ground.
See Fig. 3-1 for voltage test point
location.
d. Check—for a
meter reading of +175 to +247.5 volts.
NOTE
Ripple
and regulation of the
individual supplies can
be
checked
using the
procedure given under
Troubleshooting
Techniques
in the Maintenance
section of
this
manual.
4.
Check
Calibrator
Output Voltage
a.
Connect
the digital voltmeter between the
CALIBRATOR current loop
on the front panel, and a
ground
test point.
See Fig. 3-1 for ground test point
location.
R85t>
30
V ADJ
o
'¾:
*
GROUND
Fig.
3-1. Locations of power-supply test
points
and
adjustments.
so
2I34O7
3-3
Summary of Contents for 5110
Page 6: ...5110 Fig 1 1 5110 Oscilloscope 2134 01 ...
Page 22: ......
Page 34: ......
Page 42: ......
Page 83: ...5110 I43XII3 2134 15 ...
Page 84: ... i 143 I132 2134 15 BLOCK DIAGRAM ...
Page 85: ... t 2 E I 5110 S LNHNOdWOD BOVJHaiNI ...
Page 88: ...J4 O3 INTERFACE J INTERFACE A NOV 1978 ...
Page 92: ...AUXILIARY BOARD 1143 2134 18 AUXILIARY BOARD ...
Page 95: ...DEFLECTION AMPLIFIERs ...
Page 102: ...LV POWERSUPPLY CALIB COMPONENTS A 4 L V Power Supply circuit board 5110 ...
Page 107: ...SIGNALS OUT OPTION 7 ONLY SI 10 OPT 7 213 26 REV A FE BI979 SIGNALS OUT COPTION T ONLY n ...
Page 114: ......
Page 115: ...5110 OSCILLOSCOPE ...
Page 116: ...FIG 2 MAINFRAME p 5110 OSCILLOSCOPE ...
Page 117: ...1 ...
Page 121: ...o ...
Page 122: ...S 1 5110 OSCILLOSCOPE ...
Page 123: ...13 FIG 4 RACKMOUNT CABINET 5110 OSCILLOSCOPE ...
Page 124: ......