Adjustment-5110
Table
3-1
(cont)
LIST
OF TEST EQUIPMENT REQUIREMENTS FOR
ADJUSTMENT
Description
Performance
Requirements
Application
Example
Coaxial
cable
Impedance,
50 Ω, length,
18 inch; connectors,
bnc.
Step
13.
a. TEKTRONIX part
012-0076-00.
Termination
Impedance, 50 Ω; accuracy,
within
2%, connectors,
bnc.
Step
13.
a. TEKTRONIX part
011-0049-01.
Tee
connector
Connectors, bnc.
Step 13.
a. TEKTRONIX
part
103-0030-00.
Insulated
Screwdriver
Length,
1
1/2-inch shaft
or longer;
plastic shaft
and
handle with metal
screwdriver
tip.
Recommended for
all
adjustments.
a.
TEKTRONIX part
003-0000-00.
PRELIMINARY
PROCEDURE FOR
ADJUSTMENT
NOTE
The
oscilloscope must
be adjusted within an am
bient
temperature range of
+20° C to +30° C
for best
overall
accuracy and to meet the electrical
characteristic tolerances
given as Performance Re
quirements
in the Specification section of this
manual.
Information
given as Supplemental Infor
mation in the Specification
section is provided for
user
information only, and should not be interpreted
as
Performance
Requirements.
1.
Remove
the cabinet sides and
bottom from
the
oscilloscope
(refer
to Cabinet Removal in the
Maintenance
section
of this manual).
2.
Check the
rear panel of the oscilloscope to ensure
that the indicated line voltage and the line voltage source
are
the
same
(refer to Operating Voltage in
the
Maintenance section of this manual).
3.
Ensure
that all test equipment is suitably
adapted
to
the line voltage to be
applied.
4. If applicable,
install the TM 500-series test equip
ment
into
the test equipment power module.
5.
Install a
vertical
amplifier unit into the left vertical
compartment
of the oscilloscope.
6.
Install a time-base unit into the horizontal
compart
ment
of
the oscilloscope.
7.
Connect
the equipment
under test and the test
equipment
to a suitable line voltage source. Turn all
equipment
on and
allow at least
20 minutes for the
equipment to stabilize.
NOTE
Titles
for external controls of the oscilloscope are
capitalized
in this procedure
(e.g. INTENSITY,
POWER). Internal
adjustments are initial capitalized
only (e.g. Intensity Range,
Vertical Gain).
Initial
Control
Settings
Set
the
following controls
during warm-up
time:
OSCILLOSCOPE
INTENSITY,
FOCUS
Set for well defined
trace
and
normal brightness.
AMPLIFIER
PLUG-IN
Display
On
Position
Centered
Voits/Div
1
Volts/Div
Cal
Fully
clockwise
Input
coupling
de
3-2
Summary of Contents for 5110
Page 6: ...5110 Fig 1 1 5110 Oscilloscope 2134 01 ...
Page 22: ......
Page 34: ......
Page 42: ......
Page 83: ...5110 I43XII3 2134 15 ...
Page 84: ... i 143 I132 2134 15 BLOCK DIAGRAM ...
Page 85: ... t 2 E I 5110 S LNHNOdWOD BOVJHaiNI ...
Page 88: ...J4 O3 INTERFACE J INTERFACE A NOV 1978 ...
Page 92: ...AUXILIARY BOARD 1143 2134 18 AUXILIARY BOARD ...
Page 95: ...DEFLECTION AMPLIFIERs ...
Page 102: ...LV POWERSUPPLY CALIB COMPONENTS A 4 L V Power Supply circuit board 5110 ...
Page 107: ...SIGNALS OUT OPTION 7 ONLY SI 10 OPT 7 213 26 REV A FE BI979 SIGNALS OUT COPTION T ONLY n ...
Page 114: ......
Page 115: ...5110 OSCILLOSCOPE ...
Page 116: ...FIG 2 MAINFRAME p 5110 OSCILLOSCOPE ...
Page 117: ...1 ...
Page 121: ...o ...
Page 122: ...S 1 5110 OSCILLOSCOPE ...
Page 123: ...13 FIG 4 RACKMOUNT CABINET 5110 OSCILLOSCOPE ...
Page 124: ......